Upgrade Converter Firmware - Xylem MJK SuSix Installation And User Manual

Turbidity and suspended solids transmitter
Table of Contents

Advertisement

Turbidity and Suspended Solids Transmitter
D. Upgrade Converter Firmware
1. Select "Firmware" in the menu bar and click "Converter".
2. Locate the "MJK Field FW" directory on the CD-ROM, select the
"842xxx-xxx.hex" file (or similar) and click "Open".
3. Select the required SuSix
"Converter Firmware" dialogue box appears and prompts for selecting
the converter to program.
The converter firmware is then transferred, and the converter Flash
programming is initiated. Expect this process to take a few minutes.
4. When the Flash programming is done, click OK on the PC and the
Display Unit to return to the SuSix
SuSix
®
5. If required you may adjust the SuSix
pressing the two outmost keys simultaneously, press the up/down
keys as required, and save the new setting by again pressing the two
outmost keys simultaneously.
EN 5.10 SuSix Manual 2007
(option 1, 2, 3 or 4) and click OK when the
®
®
Display Unit screen contrast by
®
104
display.
FW: 841013-005/843003-001