Turbidity and Suspended Solids Transmitter
Analogue Output
The analogue output is an active output with a maximum load of 800Ω.
The analogue output can be programmed for indication of primary units.
See details on page 61.
Digital Output
SuSix
has two digital outputs - DO1 with an opto (light triggered) relay
®
and DO2 with an electro-mechanical relay.
They can both be programmed for the following functions:
-
Sensor error
-
High and low alarms
DO1 shares the common terminal (Com) with DI.
See also application examples in Appendix E. Digital Input/Output Con-
nections on page 111.
EN 5.10 SuSix Manual 2007
Analogue Output
Terminal
AO
+
AO
-
Opto Relay
Terminal
DO1
Com
DO1
NO
Electro-mechanical relay
Terminal
DO2
Com
DO2
NO
34
Designation
4-20 mA
4-20 mA
Designation
Max. 50 V DC /120 mA
Designation
Max. 50 V DC /1 A
FW: 841013-005/843003-001