Dissolved Oxygen Transmitter

mA Output

When a Oxix® is connected to a power supply for the first time, the mA out-
put is automatically set to provide 4 mA at zero and 20 mA at the theoretical
maximum value.
Changes in the mA setting will not affect the relay output settings.
Note:
Both values can be set in the range 10 % to 120 %, making it
possible to increase or decrease the mA signal.
This menu is for display only.
Units
D.O. Units
%
mg/l
ppm
Back
Select
Temperature
Temperature unit
°C
°F
Back
Select
The mA output is an active output, and the maximum load is 800 Ohm.
The upper limit for the mA output is 20,5 mA
3,75 mA indicates the mA output is not in use
3,5 mA indicates values under 4 mA level
20,5 mA indicates value higher than 20 mA level.
GB Oxix Manual 1209
Enable mA output?
Yes
No
Back
OK
4mA output
4 mA =
Back
OK
20 mA output
20 mA =
Back
mA output
Select
OK
Yes
00.00 mg/l
Select
OK
25.00 mg/l
Select
OK
53
Cleaning
Electro-mechanical relay
Opto relay
Not in use
Back
Select
Interval
Use DI to clean
Clean now
Maintenance
Back
Select
Interval
Interval
Off
On
Back
Select
Select cleaning interval
(20 - 255 Min)
Back
Select
Cleaning time
005 sec
Back
Select
FW: 841013-009, 846001-002, 841513-009
Measurement delay
10 sec
OK
OK
Use
No
Yes
OK
Bac
Dela
060 Min
>
Bac
DI fu
NO
NC
OK
Bac
Clea