Wiring Schematic #2 - Xylem MJK SuSix Installation And User Manual

Turbidity and suspended solids transmitter
Table of Contents

Advertisement

Turbidity and Suspended Solids Transmitter

Wiring Schematic #2

for "SuSix
with Blind Lid and a Magflux
®
common wall mounted Display Unit" on the opposite page (page 40).
SuSix
with Blind Lid
®
2-wire, shielded twisted cable
1. Connect lead "1" to slot "Display A"
2. Connect lead "2" to slot "Display B"
3. Connect the twisted shield to slot
"Display -"
Magflux
with Blind Lid
®
2-wire, shielded twisted cable
1. Connect lead "1" to slot "Display A"
2. Connect lead "2" to slot "Display B"
3. Connect the twisted shield to slot
"Display -".
4-wire, shielded twisted cable
1. Connect lead "1" to slot "Display +"
2. Connect lead "2" to slot "Display A"
3. Connect lead "3" to slot "Display B"
4. Connect the twisted shield to slot
"Display -".
Wall Mounted Display Unit
4-wire, shielded twisted cable
1. Connect lead "1" to slot "Display +"
2. Connect lead "2" to slot "Display A"
3. Connect lead "3" to slot "Display B"
4. Connect the twisted shield to slot
"Display -".
EN 5.10 SuSix Manual 2007
®
207920
A
1
2
3
4
5
6
7
B
1
2
3
4
5
6
7
42
with Blind Lid connected to a
+DC GND A
+
8
9
1
8
9
1
FW: 841013-005/843003-001
Max. four MJK co
can be connected
Converter #1
B
A
B
-
+
-
C
NO
C1
-Display-
-mA-
-DI-
-D01-
1
2 Shield
1
2 Shield
2
3
4 and
shield
2
3
2 wire
shielded ca
N01
C2 N02
-D02-

Advertisement

Table of Contents
loading

Table of Contents