Turbidity and Suspended Solids Transmitter

Install the new gasket and wiper

1. Move the sensor from the process fluid.
2. Disconnect the sensor from the converter and all electrical connec-
tions.
3. Clean and wipe the sensor off thoroughly.
4. Position the probe in a stable,
upright position.
5. Unscrew by hand the wiper with shaft counter-clockwise
(see figure 1).
DO NOT loosen or remove the screw!
6. Turn by hand the replacement wiper with shaft and O-rings
(figure 2) clockwise (figure 3).
7. Tighten with a torque of 15 Ncm.
8. It is very important that the wiper is secured in its resting position indi-
cated by the white area
(see figure 4). Otherwise the wiper will block the optics and the sensor
will not measure correctly.
9. If the wiper is not secured within the white area, loosen the screw and
reposition the wiper correctly and tighten the screw again.
10. Reconnect the sensor to the converter and put the sensor back into
operation.