Hydronix Hydro-Control VI Operator's Manual page 44

Hide thumbs Also See for Hydro-Control VI:
Table of Contents

Advertisement

Using Moisture Control
The CALC Mode calibration procedure
2.8
Task
Identify a good batch to use as a basis for the calibration.
Press the Menu button.
Press the Mix Log button.
Select the mix in the Mix Log that corresponds to the
selected good batch. Check the stability of the mix by
setting the mix log to Display Unscaled and looking at the
Dry and Wet Mix Deviation. These should be less than 3
and preferably less than 0.5. For sensors with multiple
measurement modes check the deviation for each available
mode.
Press View Mix Trace.
Press the Calibrate button.
Choose which calibration mode is most appropriate for the
mix and press OK
44
Hydro-Control VI Operators Guide HD0456 Rev 1.7.1
Action
Select a mix that has the following:
Correct final moisture
Stable
dry
readings (low deviation
values)
More than 5 unscaled
between dry and wet
average readings
Press
Press
Select the mix in the Mix Log
Press
Press
Chapter 8
and
wet