Medical equipment needs special precautions regarding EMC and need to be
installed and put into service
according to the EMC information provided in this section.
Guidance and manufacturer's declaration ± electromagnetic immunity
The FreeStyle 5 is intended for use in the electromagnetic environment specified below. The
customer or the user of the FreeStyle 5 should assure that it is used in such an environment.
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions and
voltage variations
on power.
IEC 61000-4-11
Power frequency
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
MN167-1 rev. D
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 1 kV line to line
± 2 kV line to
earth
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5s
3 A/m
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
Not Applicable
± 1 kV line to line
± 2 kV line to
earth
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 s
3 A/m
Electromagnetic environment ±
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
FreeStyle 5 requires continued
operation during power mains
interruptions, it is recommended
that the FreeStyle 5 be powered
from an uninterruptible power
supply (UPS) or a battery.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital environment.
EN
EN-40