CAIRE NewLife Intensity 10 User Manual
Hide thumbs Also See for NewLife Intensity 10:

Advertisement

Quick Links

NewLife
Intensity 10
®
User Manual
(English)

Advertisement

Table of Contents
loading

Summary of Contents for CAIRE NewLife Intensity 10

  • Page 1 NewLife Intensity 10 ® User Manual (English)
  • Page 2: User Controls & System Status Indicators

    NewLife Intensity 10 User Controls & System Status Indicators ISO 7000; Graphical symbols for use on Council Directive 93/42/EEC; concerning equipment—Index and synopsis medical devices Authorized representative in the Read user’s manual before operation. European Community Reg. # 1641 If the product unique device identifier...
  • Page 3: Why Your Physician Prescribed Oxygen

    NewLife Intensity 10 NewLife Oxygen ® IEC 60601-1: Medical electrical equipment Part 1 Concentrator General requirements for basic safety and essen- tial performance This user manual will acquaint you with the Drip Proof Equipment - IP21 NewLife oxygen concentrator. Make sure you read...
  • Page 4: Operator Profile

    NewLife Intensity 10 Operator Profile WARNING: “NO SMOKING – OXYGEN IN USE” SIGNS MUST BE PROMINENTLY Concentrators are intended to supply supplemental DISPLAYED IN THE HOME, OR WHERE oxygen to users suffering from discomfort due to OXYGEN IS IN USE. USERS AND THEIR ailments which effect the efficiency of one’s lungs...
  • Page 5 NewLife Intensity 10 Getting to Know Your NewLife Oxygen Concentrator First, become familiar with the important parts of your NewLife Oxygen Concentrator (Figures 1a, 1b). A. On/Off (I/0) Power Switch: Starts and stops the operation of the unit. B. Circuit Breaker Reset Button: Resets the unit after electrical overload shutdown C.
  • Page 6: Intended Use

    NewLife Intensity 10 Safety Guidelines Indications for Use Intended Use WARNING: CAREFULLY REVIEW AND The CAIRE NewLife Oxygen Concentrator is FAMILIARIZE YOURSELF WITH THE intended for the administration of supplemental ox- FOLLOWING IMPORTANT SAFETY INFORMATION ABOUT THE NEWLIFE ygen. The device is not intended for life support nor INTENSITY OXYGEN CONCENTRATOR.
  • Page 7 NewLife Intensity 10 WARNING: THE OXYGEN CONCENTRATOR CAUTION: Federal (USA) law restricts SHOULD NOT BE USED ADJACENT TO OR this device to sale or rental by order of a STACKED WITH OTHER EQUIPMENT. IF ADJA- physician or other licensed health care CENT OR STACKED USE IS UNAVOIDABLE, THE provider.
  • Page 8 A firebreak is required for use with any cannula. Note: If the unit has not been used for an extended • CAIRE offers a firebreak intended to be used in con- period of time, it needs to operate for several minutes junction with the oxygen concentrator.
  • Page 9: Specifications

    NewLife Intensity 10 Specifications NewLife Intensity 10 Flow Rates* 2–10 LPM ±10% of indicated setting, or 200 mL, whichever is greater** Dimensions 27.5 x 16.5 x 14.5 in (69.9 x 41.9 x 36.8 cm) Weight 58 lbs (26.3 kg) Sound Pressure...
  • Page 10 NewLife Intensity 10 Operating Instructions 1. Locate the unit near an electrical outlet in the 8. Set the flowmeter adjustment knob to the room where you spend most of your time. prescribed LPM. The concentrator is now ready for use.
  • Page 11: Proper Setting Of Oxygen Flowmeter

    NewLife Intensity 10 WARNING: THIS DEVICE SUPPLIES CAUTION: Always operate the unit in an HIGH-CONCENTRATION OXYGEN THAT upright position. PROMOTES RAPID BURNING. DO NOT ALLOW SMOKING OR OPEN FLAMES Proper Setting of Oxygen Flowmeter WITHIN THE SAME ROOM OF (1) THIS...
  • Page 12 NewLife Intensity 10 Filters Operating With Humidifier Air enters the NewLife unit through an air intake Following these steps if your physician prescribed gross particle filter located on the back off the oxy- an oxygen humidifier as part of your therapy: gen concentrator.
  • Page 13: Nasal Cannula

    NewLife Intensity 10 Note: The use of some oxygen administration acces- Note: To Equipment Provider: The following oxygen sories not specified for use with this oxygen concen- administration accessories are recommended for use trator may impair its performance. Recommended with the NewLife Oxygen Concentrator: accessories are reference within this manual.
  • Page 14: Safety Features

    NewLife Intensity 10 Safety Features Oxygen Monitor The following information will acquaint you with The oxygen monitor is a small electronic device safety features of the NewLife Oxygen Concen- within the NewLife Oxygen Concentrator that trator. Make sure you read and understand all the...
  • Page 15 NewLife Intensity 10 Operating Instructions—Dual Flow The NewLife Intensity 10 unit’s 10-liter dual flow option allows a single concentrator to meet the high flow requirements of a 10 lpm patient or the needs of two patients, in any combination of flows up to 10 lpm.
  • Page 16: Cleaning The Nebulizer

    NewLife Intensity 10 3. Repeat this procedure until the prescribed amount of medication nebulizes or the Prescribed amount of treatment time elapses (whichever occurs first). 4. If your physician or health care professional instructed you to take short rest periods During your treatment, make sure you turn the air valve to the OFF position.
  • Page 17: Cleaning, Care, And Proper Maintenance

    NewLife Intensity 10 Materials in Direct or Indirect Contact with Operator NewLife Intensity 10: Concentrator casing ..Valtra/ABS/Polystyrene Mains cable ............PVC Dust filter ..........Polyester ON/OFF switch ........ Thermoplastic Casters ............Nylon Flow adjustment .......ABS/Polycarbonate Gas outlet .......Chrome Plated Brass Printed labels ..........Lexan...
  • Page 18: Reserve Oxygen Supply

    NewLife Intensity 10 Reserve Oxygen Supply WARNING: USE ONLY WATER-BASED LOTIONS Your Equipment Provider may recommend another OR SALVES THAT ARE OXYGEN COMPATIBLE source of supplemental oxygen therapy in case PRIOR TO OR DURING OXYGEN THERAPY. NEVER there is a mechanical failure or a power outage.
  • Page 19: Alarm Conditions

    NewLife Intensity 10 Alarm Conditions All alarms are low priority alarms. Alarm Indicates Action General malfunction yellow light High Product Tank Pres- Ensure flowmeter is open to minimum flow rate or sure OR Low Product higher. Ensure cannula is not kinked or obstructed.
  • Page 20 Intermittent alarm sounds. Equipment malfunction. Set I/0 power switch to 0 position, use your reserve oxygen supply and consult your Equip- ment Provider immediately. NewLife Intensity 10 displays Refer to Alarm Conditions Refer to Alarm Conditions table. alarm and produces intermit- table.
  • Page 21 NewLife Intensity 10 Standard Accessories country-specific power cords where noted above. Nasal Cannula with 7 feet (2.1 m) CU002-1 Contact CAIRE or your oxygen provider if alternate of tubing (6 LPM max) options are needed for order. Oxygen Outlet Adapter...
  • Page 22: Emc Testing

    NewLife Intensity 10 EMC Testing Medical equipment needs special precautions regarding EMC and needs to be installed and put into ser- vice according to the EMC information provided in this section. Guidance and Manufacturer’s Declaration —Electromagnetic Emissions The NewLife is intended for use in the electromagnetic environment specified below. The customer or the user of the NewLife should assure that it is used in such an environment.
  • Page 23: Guidance And Manufacturer's Declaration - Electromagnetic Immunity

    NewLife Intensity 10 Guidance and Manufacturer’s Declaration± Electromagnetic Immunity The NewLife is intended for use in the electromagnetic environment specified below. The customer or the user of the NewLife should assure that it is used in such an environment. Immunity test...
  • Page 24 To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the NewLife Intensity 10 is used exceeds the applicable RF compliance level above, the NewLife should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the NewLife Intensity 10.
  • Page 25 NewLife Intensity 10 Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment Test Band Service Modulation Maximum Distance Immunity frequency (MHz) Power (W) Test Level (MHz) (V/m) Pulse modulation 380-390 TETRA 400 18 Hz GMRS 460, FRS 430-470 ±5 kHZ deviations...
  • Page 26 NewLife Intensity 10 Recommended separation distances between portable and mobile RF communications equipment and the NewLife Units The NewLife is intended for use in an electromagnetic environment in which radiated RF disturbances are con- trolled. The customer or the user of the NewLife can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the NewLife as recommended below, according to the maximum output power of the communications equipment.
  • Page 27 NewLife Intensity 10 Notes PN MN239 F | User Manual ENG - 27...
  • Page 28 35619 Braunfels, Germany Tel: +49 (0) 6442-962073 E-mail: info@mps-gmbh.eu CAIRE and CAIRE Inc. are registered trademarks of CAIRE Inc. Please visit our website below for a full listing of trademarks. Trademarks: www.caireinc.com/corporate/trademarks/. © Copyright 2022 CAIRE Inc. All Rights Reserved.