  • Page 1 User Manual FreeStyle & FreeStyle ™ ™...
  • Page 2: Table Of Contents

    Contents User Controls & System Status Indicators ..........3 What is the Oxygen Concentrator ............4 Why Your Physician Prescribed Oxygen ..........4 ..................5 Unpacking Your FreeStyle or FreeStyle 5 ..........5 Getting to Know Your FreeStyle or FreeStyle 5 ........6 Indications for Use .................8 Contraindications for Use ..............8 Safety Guidelines ...................8 Operating Instructions ................13...
  • Page 3: User Controls & System Status Indicators

    User Controls & System Status Indicators Symbol Definition Symbol Definition Consult instructions for use Do not disassemble No Smoking Icon: Do not smoke near unit. Use no oil or grease. This side up No open or naked flames. RTCA/DO-160 Section 21 RTCA DO160 Section 21 Category M Safety agency for CAN/CSA C22.2 No.
  • Page 4: What Is The Oxygen Concentrator

    What is the Oxygen Concentrator The air we breathe contains approximately 21% oxygen, 78% nitrogen, and 1% other gasses. In the the user breathes in and delivers a pulse of oxygen FreeStyle and FreeStyle 5, room air is drawn into the machine through the air intake.
  • Page 5: Unpacking Your Freestyle Or Freestyle 5

    Operator Profile Concentrators are intended to supply supplemental Although oxygen therapy can be prescribed for oxygen to users suffering from discomfort due to users of all ages, the typical oxygen therapy user is older than 65 years of age and suffers from a to transfer oxygen in the air to their bloodstream.
  • Page 6: Getting To Know Your Freestyle Or Freestyle 5

    Getting to Know Your FreeStyle or FreeStyle 5 Become familiar with the key features of the FreeStyle and FreeStyle 5 and their control panels Handle Control Panel Oxygen Outlet Air Outlet(s) Inlet for Power Connections Battery Pack Air Intake/Filter Battery Pack Cover Release Button 6 —...
  • Page 7 Alarm Indicator Light Battery Gauge Indicator Lights Pulse Flow Indicator Battery Status Button Flow Selection Buttons Flow Selection Buttons: Battery Status Button: To power the device ON or OFF, press the numeric Press once to display the battery gauge indicator to determine how much time is remaining on the charge of the installed battery pack.
  • Page 8: Indications For Use

    Important! Contraindications Safety Instructions are de ned as for Use follows: WARNING: IN CERTAIN CIRCUMSTANCES, WARNING: IMPORTANT SAFETY THE USE OF NON-PRESCRIBED OXYGEN INFORMATION FOR HAZARDS THAT MIGHT CAN BE HAZARDOUS. THIS DEVICE SHOULD CAUSE SERIOUS INJURY. ONLY BE USED WHEN PRESCRIBED BY A PHYSICIAN.
  • Page 9 WARNING: IF THE OXYGEN WARNING: DO NOT ALLOW EITHER THE AIR CONCENTRATOR HAS BEEN INTAKE OR THE AIR OUTLET EXHAUST TO DROPPED, DAMAGED OR EXPOSED BECOME BLOCKED. DO NOT DROP OR INSERT TO WATER PLEASE CONTACT ANY OBJECTS INTO ANY OPENINGS ON YOUR HOME CARE PROVIDER FOR THE DEVICE.
  • Page 10 CAUTION: In the event of an alarm or WARNING: “NO SMOKING – OXYGEN IN USE” you observe the Oxygen Concentrator SIGNS MUST BE PROMINENTLY DISPLAYED is not working properly; consult the IN THE HOME, OR WHERE OXYGEN IS IN USE. Troubleshooting section in this manual.
  • Page 11 CAUTION: Always disconnect AC power supply from the wall before disconnection Note: Ensure the cannula is fully inserted and secure. the AC power supply from the oxygen This ensures that the Oxygen Concentrator can properly concentrator. detect inspiration for oxygen delivery. During inhalation, CAUTION: Do not position the Concentrator so that it you should hear or feel oxygen flow to the prongs of the is difficult to access the power cord.
  • Page 12 FreeStyle FreeStyle 5 Dimensions 8.6 in x 6.1 in x 3.6 in 10.7 in x 6.6 in x 4.4 in (21.8 cm x 15.5 cm x 9.1 cm) (27.2 cm x 16.8 cm x 11.2 cm) Weight (w/ Battery 4.9 lbs (2.2 kg) 6.7 lbs (3.0 kg) Pack) Nominal Sound Level...
  • Page 13: Operating Instructions

    Operating Instructions 1 Locate and position the FreeStyle or FreeStyle 5 3 Insert the AC or DC power supply into the in a proper location for use. receptacle on the right side of the unit if recharging the battery pack and/or running from that they air intake and air outlets are not be the external power source.
  • Page 14 5 Turn on the FreeStyle or FreeStyle 5 by pressing on the device. A brief audible alarm will sound to indicate the device has powered on and the green light above remain illuminated. 6 Position the nasal cannula on your face per the cannula and breathe normally.
  • Page 15: Power Supplies

    DISCONNECTED FROM EXTERNAL WARNING: USE ONLY THE POWER POWER BEFORE REMOVING THE SUPPLIES OUTLINED IN THIS USER’S BATTERY PACK. MANUAL. USE OF NON-CAIRE POWER SUPPLIES AND BATTERY PACKS CAN CAUSE DAMAGE TO THE UNIT AND IMPROPER OPERATION. Battery Pack The FreeStyle or FreeStyle 5 can be operated by the rechargeable battery pack which is supplied with the device.
  • Page 16 BATTERY OPERATING DURATION: To check the battery charge when it is not installed in the FreeStyle or FreeStyle 5, turn the battery on When external AC or DC power is disconnected, its back and press the TEST button. The battery the concentrator will automatically switch over to using battery power if the battery pack is installed the right of the TEST button to indicate the level of...
  • Page 17: Freestyle Ac & Dc Power Supply

    FreeStyle AC & DC Power Supply To connect the FreeStyle to AC Power: power supply to keep your concentrator powered 1. Connect the universal power supply to the wherever you go. It replaces separate AC and DC power connector inlet on the side of the power supplies so you only have a carry and keep FreeStyle.
  • Page 18 Ensure the green light on top of the Universal 2. Connect the DC power cord to the inlet marked power supply is illuminated. This indicates “ACCESSORY DC IN” on the universal power that the power supply is receiving power from supply the outlet.
  • Page 19: Freestyle 5 Ac & Dc Power Supply

    FreeStyle 5 AC & DC Power Supply 2. Connect the AC power cord to the inlet on the supplies for AC and DC operation. power supply. The FreeStyle 5 includes the following: 1. AC power supply which operates on AC Power 2.
  • Page 20 4. Ensure the green light on top of the DC supply DC POWER is illuminated. This indicates that the power supply is receiving power from the outlet. If DC power is for use in the accessory outlets of the green light is not illuminated, disconnect automobiles, boat, or other motor vehicles.
  • Page 21: Cleaning, Care & Routine Maintenance

    Cleaning, Care & Routine Maintenance ROUTINE MAINTENANCE Servicing of internal components inside of the cabinet of the FreeStyle or FreeStyle 5 must be performed by the oxygen provider. The operator is responsible for the following maintenance procedures. • Weekly – Clean the Air Intake Filter WARNING: DO NOT ATTEMPT TO OPEN OR REMOVE THE PLASTIC COVERS OF THE FREESTYLE OR...
  • Page 22 CLEANING may also cause the battery to generate heat, smoke, Turn OFF the unit and disconnect from AC or DC be stored in a stable environment characterized power before any cleaning or disinfection activity. DO NOT spray the outer case directly. Use a corrosive gasses.
  • Page 23: Alarm Conditions

    Alarm Conditions START-UP ALARM face, and ensure that you are breathing through your nose. If the alarm continues to sound, change A brief alarm sounds when the FreeStyle or to another source of oxygen as available, and The concentrator begins to operate once the brief alarm stops.
  • Page 24: Troubleshooting

    Troubleshooting Problem Probable Cause Solution FreeStyle does not operate when Battery is discharged. Power the unit through the DC outlet, or an AC a pulse flow selection button is outlet. Malfunction. pressed. Contact your Equipment Provider, and change to another source of oxygen as necessary. A continuous alarm sounds FreeStyle has not detected a Check the cannula connection.
  • Page 25: Travel Information

    Air Intake Filter FI144-1 BT032-4 (UK Cord) Note: Additional options may be available for country- specific power cords where noted above. Contact CAIRE or your oxygen provider if alternate options are needed for order. FreeStyle & FreeStyle 5 User Manual — 25...
  • Page 26: Carrying Bag Instructions

    Carrying Bag Instructions SHOULDER STRAP INSTALLATION To insert and secure your FreeStyle or FreeStyle 5 unit in the carrying bag, utilize the following 1. Unclip the carry handle. instructions. INSERTING UNIT INTO CARRYING 1. Insert unit into bottom of bag, properly aligning 2.
  • Page 27: Materials In Direct Contact With The Operator

    Materials in Direct Contact with the Operator • Concentrator Casing ........................PC/ABS. • Concentrator Control Panel ................• Concentrator Handle ......................Polycarbonate • Oxygen Outlet, Nozzle ......................Polycarbonate • Unit Label ............................Lexan • Cord Connectors ...................Polycarbonate/Vinyl chloride • Power Cord ..........................PVC, Metal •...
  • Page 28 Guidance and Manufacturer’s Declaration—Electromagnetic Immunity The FreeStyle and FreeStyle 5 are intended for use in the electromagnetic environment specified below. The customer or the user of the FreeStyle and FreeStyle 5 should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level...
  • Page 29 Guidance and Manufacturer’s Declaration–Electromagnetic Immunity The FreeStyle and FreeStyle 5 are intended for use in the electromagnetic environment specified below. The customer or the user of the FreeStyle and FreeStyle should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level...
  • Page 30 Recommended separation distances between portable and mobile RF communications equipment and the FreeStyle or FreeStyle 5 The FreeStyle and FreeStyle 5 are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the FreeStyle and FreeStyle 5 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the FreeStyle and FreeStyle 5 as recommended below, according to the maximum output power of the communications equipment.
  • Page 31: Notes

    Notes FreeStyle & FreeStyle 5 User Manual — 31...
  • Page 32 Tel: +49 (0) 6442-962073 E-mail: info@mps-gmbh.eu Worldwide: www.CAIREmedical.com Copyright © 2016 Chart Industries. CAIRE Inc. reserves the right to discontinue its products, or change the prices, materials, equipment, quality, descriptions, specifications and/or processes to its products at any time without prior notice and with no further obligation or consequence.