Problem
Rapid intermittent alarm
condition, and the (red)
alarm light illuminates
continuously.
Unit does not start on
battery power.
Delay in recharging
internal battery.
Unit alarms while in
automobile and
connected to the
DC outlet.
All other problems.
MN167-1 rev. D
Probable Cause
A general malfunction
has occurred.
Unit may be hot or
cold if left outdoors
such as in an
automobile.
Internal battery
exceeds charging
temperature.
No power to the unit if
battery depleted and
DC outlet not
charging.
Solution
Change to another
source of oxygen as
available, and contact
your Equipment Provider.
Allow the unit to reach
normal operating
temperature, which may
take several minutes if
exposed to temperature
extremes. Temporarily
connect your AC or DC
power supply to the unit's
power inlet and power
source, as needed, to
reset the unit's internal
battery.
Unit may be operated;
however, charging may
not resume until battery
temperature is reduced.
(See Battery Charger
notes.)
Disconnect the DC power
supply from the
automobile outlet, restart
the automobile, and then
reconnect the DC power
supply into the
automobile DC outlet to
reset the breaker within
DC power supply.
Change to another
source of oxygen as
available, and contact
your Equipment Provider.