CAIRE NewLife Elite Technical Manual
Hide thumbs Also See for NewLife Elite:

Advertisement

Quick Links

NewLife Family Technical Manual MN240-1
MN240-1 Rev C
1

Advertisement

Table of Contents
loading

Summary of Contents for CAIRE NewLife Elite

  • Page 1 NewLife Family Technical Manual MN240-1 MN240-1 Rev C...
  • Page 2: Table Of Contents

    Table of Contents Section 1.0 Introduction Important Notice and Symbol Explanations Equipment Provider Responsibility Functional Specifications Section 2.0 Operation Check and Oxygen Concentration Test Description of Operation Operation Check Alarm System 2.3.1 Start Up Test 2.3.2 Power Failure Alarm Test 2.3.3 General Malfunction Alarm Test 2.3.4 Low O Alarm Test...
  • Page 3 Section 5.0 Main Components Components Cabinet Removal 5.2.1 Removing Side Panel(s) 5.2.2 Removing Back Panel 5.2.3 Removing Lower Front Panel 5.2.4 Removing Control Panel 5.2.5 Superstructure 5.2.6 Caster Replacement Compressor 5.3.1 Compressor Replacement 5.3.2 Capacitor Replacement Solenoid Valves 5.4.1 Feed/Waste Valve Replacement 5.4.2 Solenoid Valve Coil Replacement Sieve Beds 5.5.1 Sieve Bed Replacement...
  • Page 4 5.13 Hour Meter Replacement 5.14 Flowmeter Replacement 5.15 Power Cord Replacement 5.16 Oxygen Monitor Circuit Board Replacement Section 6.0 Troubleshooting Operating Pressure Test 6.1.1 High Operating Pressure 6.1.2 Low Operating Pressure General Troubleshooting Troubleshooting Chart Tool Kit and Pressure Test Gauge Section 7.0 EMC TESTING 7.1 EMC Testing...
  • Page 5: Important Notice And Symbol Explanations

    1.0 Introduction 1.1 Important Notice and Symbol Explanations As you read the manual, pay special attention to the WARNING, CAUTION, and NOTE messages. They identify safety guidelines or other important information as follows: Describes a hazard or unsafe practice that can result in severe bodily injury or death.
  • Page 6 MN240-1 Rev C...
  • Page 7 MN240-1 Rev C...
  • Page 8 MN240-1 Rev C...
  • Page 9 MN240-1 Rev C...
  • Page 10 MN240-1 Rev C...
  • Page 11 MN240-1 Rev C...
  • Page 12 ▪ Inspect the condition of each NewLife unit immediately upon delivery to your business location. Note any sign of damage on the delivery receipt, and report it directly to both the, freight company and CAIRE, Inc. immediately. ▪ Check the operation of each NewLife before delivery to a patient. Confirm the oxygen concentration level is within specifications.
  • Page 13 1.3 Product Specifications 5 Liter Concentrators 8 Liter Concentrators 10 Liter Concentrators Oxygen Concentration: 1-3 LPM: 95.5% - 92.0% 2-7 LPM: 92% ± 3% 2-9 LPM: 92% ± 3% 4 LPM: 92% ± 3% 8 LPM: 90% ± 3% 10 LPM: 90% ± 3% 5 LPM: 90% ±...
  • Page 14 Product Specifications Continued Relief Pressure 45 PSIG (310 kPa) @ Compressor Monitor Yellow caution light Range: Below 82% (+ 3%) O Audible Alarm: After several minutes below specified Range Max Pressure = 60 PSIG (414kPa) Operating temperature Range: -31°F to 158°F (-35°C to 70°C) Operational Humidity: 0 to 100% Altitude Tested all models at -381 m to 3,048 m (-1,250 ft to 10,000 ft) (700-1060hPa)
  • Page 15 Product Specifications Continued 5 Liter 8 Liter 10 Liter Concentrators Concentrators Concentrators Excessive Maximum O Outlet Temperature Rise Maximum O Outlet Temperature Rise Temperatures – Outlet C (5.4 + 4.5 C (8.1 Flow Rate Range 1% maximum difference in flow rate of Within ±...
  • Page 16 Product Specifications Continued 5 Liter Concentrators 8 Liter Concentrators 10 Liter Concentrators Backpressure Flow Applied Pressure * Flow Applied Pressure * Flow Applied Pressure * Effect Test 5.00 LPM @ 0 PSIG (0 kPA) 8.00 LPM @ 0 PSIG (0 kPA) 10.00 LPM @ 0 PSIG (0 kPA) 4.75 LPM...
  • Page 17: Description Of Operation

    2.2 Operation Check CAIRE tests every NewLife Oxygen Concentrator thoroughly after manufacture. You must perform the following test to ensure that no damage occurred in shipping or handling.
  • Page 18: Alarm System

    (shutting down the compressor) ** Note: The indication for a “no flow” condition on the NewLife Elite and Dual Flow NewLife Intensity 10 models is the flow ball dropping to zero. There is no audible alarm indication for the “no flow”...
  • Page 19: Power Failure Alarm Test

    Device warning label Oxygen Monitor Label Figure 2: Device warning label and oxygen monitor label. Note this is valid for units manufactured before January 1 , 2019. Oxygen Monitor (Oxygen Monitor Unit Only): In the event the oxygen monitor detects an oxygen concentration below 82%, the oxygen monitor will illuminate.
  • Page 20: General Malfunction Alarm Test

    3. While the machine is still operating, disconnect from external power. 4. Verify that the power loss alarm LED illuminates on the front panel and audible alarm activates. The audible alarm will beep approximately once every 20 seconds. If the unit has not been used for an extended period of time, it needs to operate for several minutes to recharge the power failure alarm.
  • Page 21: Oxygen Concentration Test And Specifications

    Disconnect the oxygen analyzer, and reconnect the humidifier bottle (if used). Do not measure oxygen concentration output after the product stream passes through a humidifier bottle, or erroneous readings will result. CAIRE NewLife Concentration Specifications Model Liter Flow Specification...
  • Page 22: Instructions

    3.0 User Instructions 3.1 Instructions It is important that patients thoroughly understand how to operate the CAIRE NewLife unit. This enables proper treatment, as prescribed by a licensed health care provider/physician. If users experience any discomfort or the unit alarms, they must notify their licensed health care provider/physician immediately.
  • Page 23: Air Intake Gross Particle Filter/Gpf

    Equipment Providers need to establish and implement a protocol to check oxygen concentration. CAIRE does not require preventative maintenance on the concentrator. You do not need to perform any maintenance as long as the NewLife unit remains within specifications at the desired flow rate.
  • Page 24: Recording Maintenance

    4.1.3 Recording Maintenance As the Equipment Provider, you can record all routine maintenance and repairs performed on the NewLife unit, including hours and dates of service. Electrical shock hazard. Disconnect the power cord from the electric outlet before you clean the unit to prevent accidental electrical shock and burn hazard.
  • Page 25: Main Components

    5.0 Main Components 5.1 Components The design of the CAIRE NewLife Oxygen Concentrator allows for easy access and removal of most components. This allows you to perform repair and replacement of parts with minimal time and effort. To prevent accidental electrical shock or burn, be sure to set the unit’s...
  • Page 26: Compressor

    You can continue a patient’s therapy on the NewLife unit as long as that unit’s oxygen concentration level at the prescribed liter flow rate is within CAIRE’s specifications. Sound Level The condition of the compressor’s cup seals, bearings, and other components can result in an increased sound level.
  • Page 27: Compressor Replacement

    5.3.1 Compressor Replacement Compressor Assembly Removal To remove the compressor assembly, follow the steps below: 1. Set the unit’s ON/OFF switch to the OFF position and disconnect the power cord. 2. Remove the side and lower front panels of unit. 3.
  • Page 28: Solenoid Valves

    5.4 Solenoid Valves The NewLife uses five two-way solenoid valves: two feed, two waste, and one equalization. Each valve has an open (energized) and closed (de-energized) position. As the NewLife operates, two valves are always energized. The solenoid valves of the NewLife unit require no scheduled maintenance. If a valve becomes noisy, you can easily replace the internal valve parts.
  • Page 29: Solenoid Valve Coil Replacement

    5.4.2 Solenoid Valve Coil Replacement An ohmmeter can be a useful tool in determining the condition of a valve coil. When using and ohmmeter in the NewLife unit, valve coil readings should be as indicated below: 1. Elite 120VAC units a.
  • Page 30: Sieve Beds

    5.5 Sieve Beds The NewLife unit utilizes two sieve beds, each containing molecular sieve. The unique property of molecular sieve enables it to physically attract nitrogen when air passes through this material, thus producing highly concentrated oxygen. While one sieve bed produces high concentration oxygen, the other is purged of the nitrogen it adsorbed (collected) while it concentrates oxygen.
  • Page 31: Cabinet Fan

    5.6 Cabinet Fan The cabinet fan pulls ambient air into the NewLife unit. As this air is drawn in, the air cools internal components (including the compressor) and exits out the bottom of the unit. The cabinet fan for NewLife is located in the back of the unit. Refer to the troubleshooting chart in Section 6.0 of this manual for instances where replacement of the fan may be required.
  • Page 32: Circuit Board

    5.7 Circuit Board The solid-state printed circuit board controls the timing operation of the five solenoid valves and the alarm system functions. Consult the troubleshooting chart in Section 6.0 to determine when to replace the printed circuit board. The Printed Circuit Boards (PCBs) contain components that are sensitive to electrostatic discharge (ESD) and can damage the board if not handled properly.
  • Page 33: Product Tank Replacement

    5.8 Product Tank Replacement Product Tank Removal To remove the product tank, follow the steps below: Set the unit’s ON/OFF power switch to the OFF position and disconnect the power cord. Remove the sides, lower front, and back panels. Remove the compressor. (Refer to Section 5.3.) Remove the intake resonator, which is held in place by Velcro.
  • Page 34: Product Regulator Check And Setting

    Intensity 8 and 10 and should not require adjustment. To check for proper adjustment of the product regulator, take the following steps: Procedure A: NewLife Elite 1. Set the ON/OFF switch to the ON position. 2. Allow the unit to run for five minutes.
  • Page 35: Product Regulator Cleaning Or Rebuilding

    Procedure B: NewLife Intensity 8 and 10 Use the following procedure to set the product regulator to 20 PSIG (138 kPa): 1. Disconnect the humidifier bottle, if used, and the tubing from oxygen outlet. 2. Remove the side panels and back panel. 3.
  • Page 36: Circuit Breaker Replacement

    5.10 Circuit Breaker Replacement Circuit Breaker Removal To remove the circuit breaker, follow the steps below: 1. Set the unit’s ON/OFF switch to the OFF position and disconnect the power cord. 2. Remove the left side panel. 3. Cut the heat shrink to expose the circuit breakers leads. 4.
  • Page 37 5.13 Hour Meter Replacement HM009-2 is a dual function hour meter. It has a selection switch accessible by a pinhole in the upper right corner of the hour meter. To toggle between display modes - press in the switch once using a small paperclip. ‘TMR1’...
  • Page 38 5.15 Power Cord Replacement 1. Set the unit’s ON/OFF switch to the OFF position and disconnect the power cord. 2. Remove the side and back panels. 3. Open the plastic twist clamps. 4. Cut the tie-wraps (located near the power cord push-on connectors). 5.
  • Page 39 6.0 Troubleshooting 6.1 Operating Pressure Test Testing the operating pressure is a useful diagnostic tool when a concentrator produces low oxygen concentration and requires servicing. Units functioning normally do not require operating pressure tests. Use the following procedure to test the operating pressure of the unit: Set the unit’s ON/OFF power switch to the OFF position and disconnect the power cord.
  • Page 40 6.1.2 Low Operating Pressure If the oxygen concentration level at the desired liter flow is within CAIRE’s specifications, no service is required, even if operating pressure is low. If concentration is below specifications at the desired liter flow rate with low operating pressure, this may indicate any of the following: •...
  • Page 41 6.3 Troubleshooting Chart Problem Probably Cause Solution Compressor does not run. Power No power to unit from electrical Check or restore power to outlet. outlet. failure light ( ) and intermittent audible alarm. Unit circuit breaker tripped or Reset or replace circuit breaker. faulty.
  • Page 42 Problem Probable Cause Solution Compressor relief valve releases Faulty electrical connection at Repair electrical connection. (popping sound). waste valve. Faulty solenoid valve coil. Replace valve coil. Refer to section 5.4.2 for Use of an ohmmeter can be used acceptable ohm rating. to easily determine a faulty coil.
  • Page 43 Problem Probable Cause Solution Limited or low flow Restriction in humidifier/tubing. Replace humidifier or tubing. Product regulator set too low. Adjust regulator setting. Leak. Leak test and repair leak. Reduced air intake (suction). Check compressor intake path for obstruction. Clean/remove obstruction.
  • Page 44 Pressure test adapter: (KI057-1). Outlet pressure test kit: (TU131-1 for Intensity 10 only) (CAIRE p/n: TU131-2 for Intensity 8 only) A pressure test gauge (KI036-1) to read operating system pressures and outlet pressures on the NewLife unit should always be kept available.
  • Page 45 EMC Testing 7.1 EMC Testing Tables Medical equipment needs special precautions regarding EMC and needs to be installed an put into service according to EMC information provided in this Section. MN240-1 Rev C...
  • Page 46 MN240-1 Rev C...
  • Page 47 MN240-1 Rev C...
  • Page 48 MN240-1 Rev C...
  • Page 49 Ball Ground, GA 30107 www.cairemedical.com Copyright © 2019 CAIRE Inc. CAIRE Inc. reserves the right to discontinue its products, or change the prices, materials, equipment, quality, descriptions, specifications and/or processes to its products at any time without prior notice and with no further obligation or consequence. All rights not expressly stated herein are reserved by us, as applicable.

This manual is also suitable for:

Newlife intensity 8Newlife intensity 10

Table of Contents