Do you have a question about the Eclipse 5 with autoSAT and is the answer not in the manual?
Questions and answers
David C Morandy
February 4, 2025
I have a yellow light continusly on. I **** getting alarm code 8
1 comments:
Mr. Anderson
May 14, 2025
A continuously yellow light indicates a warning or caution, such as an oxygen concentration below 85% ± 3%, on the CAIRE 5 with autoSAT. The provided context does not specify the meaning of alarm code 8.
TABLE OF CONTENTS Power Cartridge ..........19 General Information ..........4 Typical New Power Cartridge Operation Time Warning and Caution Statements ..... 4 ..............20 Introduction to the Eclipse 5 Oxygen System ... 6 Charging Algorithm ........22 Eclipse 5 Oxygen System Specifications ..7 Training The Patient ...........
Page 3
Optional Accessories ........82 Procedure - Preferred Method ....47 Purity and Flow Rate Test Procedure- Alternate Method ........47 CAIRE Inc. Customer Service Contact Assembly and Alarm Verification Tests ..48 Information ............ 83 Record Hours of Operation & Software Version ............
GENERAL INFORMATION This technical manual will familiarize you with Provider-specific information regarding the Eclipse 5 Oxygen System. Instructions in this manual are intended to help ensure that: - Providers are familiar with Eclipse 5 system components and system principles of operation - Providers are given proper guidance in the use of the Eclipse 5 and its accessories that can be conveyed to patients - Providers are made aware of the care, diagnostics, maintenance, and repair of the Eclipse 5...
Page 5
Portable humidity range is 15 to 95%. Aircraft Use: Base humidity range is 30 to 75%. CSA Certification Reg. # 2620 CAIRE confirms that this machine meets Certified for both the U.S. the Federal Aviation Administration Atmospheric pressure limitation. Reg. and Canadian markets, to the...
INTRODUCTION TO THE ECLIPSE 5 OXYGEN SYSTEM ECLIPSE 5 AC POWER SUPPLY WITH NEMA POWER CORD PN: 5941-SEQ (three LED model) PN: 20852326S (with IEC type A cord), 20852325S UNIVERSAL CART (with IEC type C cord), and 20852324S (with IEC type G cord) (two LED model) DC POWER SUPPLY PN: 5942-SEQ...
Back-Up Alarm Power 9V Internal Battery Filters Air Inlet, HEPA, Compressor Intake Device Classification IEC Class I, Type B Applied Part, IP0 (3 LED model) IEC Class II, Type BF Applied Part, IP22 (2 LED model) Pulse Dose Mode Specifications Pulse Settings 1.0 to 6.0: 16mL, 24mL, 32mL, 40mL, 48mL, 56mL, 64mL, 72mL, 80mL, 88mL, 96mL;...
Connect the equipment into an outlet on a circuit different from that which the other device(s) are connected. • Consult with CAIRE’s Technical Support Department for help. Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this manual.
Page 11
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity The Eclipse 5 is intended for use in the electromagnetic environment specified below. The customer or the user of the Eclipse 5 should assure that it is used in such an environment. Electromagnetic environment – guidance Immunity test Compliance level Electrostatic discharge (ESD) IED...
How The Eclipse 5 Works INTRODUCTION The Eclipse 5, Personal Ambulatory Oxygen System with autoSAT Technology is a portable medical device used to extract oxygen from the atmosphere, concentrate it to 87–95.6% and present the oxygen to the patient. The device will operate in Continuous Flow Mode or Pulse Dose Mode.
COMPRESSOR AND COMPRESSOR ENCLOSURE The Eclipse 5 Compressor is a two-cylinder, variable speed wobble piston compressor, driven by a highly efficient Brushless DC (BLDC) motor. When air flows into the Compressor enclosure, it passes through an air intake filter/muffler that muffles sound and filters out impurities.
CONTROL PANEL The control panel provides a user interface consisting of a membrane panel keyboard, Liquid Crystal Display (LCD), external power present indicator, Power Cartridge capacity indicator, alarm status indicators, and an audio transducer. The user interface informs the user of the system status and allows the user to set the desired flow rate and flow mode.
CONTINUOUS FLOW MODE Continuous Flow Mode delivers a constant flow of oxygen to a patient by means of tubing and a nasal cannula at rates between 0.5 LPM and 3.0 LPM. Within the Eclipse 5, concentrated oxygen is stored in a 500ml product tank at pressures in the range of 5 to 9 psi.
PROVIDER MODE FUNCTIONS Providers can access provider mode software functions using the Control Panel. All provider mode information is displayed on the LCD. The software shall advance the following Eclipse 5 display mode when the “No Smoking” icon is pressed: •...
EDAT is the world’s first computer-based data retrieval service tool for oxygen concentrators. EDAT is proprietary to CAIRE Inc., and is comprised of both hardware and software components. EDAT is CAIRE Inc.’s global solution for your service and support needs.
POWER SUPPLIES The Eclipse 5 may operate from either the AC or DC Power Supply or the Power Cartridge. When power is available from an external supply, the Eclipse 5 will draw from the external source rather than depleting the Power Cartridge.
Figure 6: Eclipse 5 DC Power Supply Located near the output cord, there is a green LED that is illuminated when the DC Power Supply is supplying 26VDC power. If the LED is not illuminated, there is no input power available. The Power Supply contains protection circuits for output over- current, input over-voltage, and internal over-temperature conditions.
The Eclipse 5 Power Cartridge operation time may be affected by several factors such as bolus size, breathing rates, ambient temperature, age of power cartridge and use over time. The table below describes the typical operating time for a new Power Cartridge.
Need help?
Do you have a question about the Eclipse 5 with autoSAT and is the answer not in the manual?
Questions and answers
I have a yellow light continusly on. I **** getting alarm code 8
A continuously yellow light indicates a warning or caution, such as an oxygen concentration below 85% ± 3%, on the CAIRE 5 with autoSAT. The provided context does not specify the meaning of alarm code 8.
This answer is automatically generated