CAIRE AirSep FreeStyle 5 Patient Manual page 217

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 197
FreeStyle 5™
The FreeStyle 5 AirBelt will completely recharge from its fully depleted state
in approximately 3 hours.
While charging a fully discharged battery, the LED will continue to blink until
25% capacity is reached. The LED will then turn solid.
Each of the four LEDs, 25% -100%, will blink as stated above, then turn solid
when the battery reaches it's capacity.
When all LEDs illuminate solid, the battery is fully charged and the LEDs will
remain solid for a period of time, then all four LEDs will turn off.
Lithium batteries may permanently lose capacity when exposed to
extremely hot temperatures with the batteries fully charged or
completely depleted. For extended storage, it is recommended
that batteries be charged 25 to 50% and remain within a
temperature range of 73° F (23° C) +/- 2°C.
Depending upon the temperature of the AirBelt, it can take
several minutes for the charging cycle to start after connecting
to power. This is a normal condition and is intended for safe
charging.
AirBelt does not need to be fully discharged before recharging.
It is recommended to charge it after each use.
MN167-1 rev. D
Figure 9: AirBelt Battery Charging set-up
AirBelt
Connecting the AirBelt
cord to the power supply
cord
PW009-2 AC/DC Power
Supply
AirBelt AC/DC
Power Supply
EN
EN-19

Advertisement

Table of Contents
loading

Table of Contents