714 Specifications; Firmware Earlier Than V2.0 - Fluke 71X Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71X Series:
Table of Contents

Advertisement

Temperature Coefficient
0.005 % of ohms range per °C for temperature ranges –10 °C to 18 °C and 28 °C to
55 °C. Ohms ranges are 400 Ω, 1.5 kΩ, and 4.0 kΩ.

714 Specifications

714 Calibrator specifications vary based on the version of the instrument. To display the
firmware version for your instrument, start with the 714 off, push and hold , then push
the power button. Find the section heading below for the displayed version and use the
specification tables in that section to test and calibrate the instrument.

Firmware Earlier than v2.0

Table 11. 714 Temperature Measure and Thermocouple Simulate
TC Type
0.1 °C
J, K, T, E, L, U
1 °C
B, R, S
Maximum input voltage: 30 V
Temperature Coefficient: 0.05 x specified accuracy per °C for temperature ranges – 10 °C to 18 °C and
28 °C to 55 °C
Range
-10 mV to 75 mV
Maximum input voltage: 30 V
Resolution
±(0.3 °C + 10 μV)
±(0.3 °C + 10 μV)
Table 12. 714 Millivolt Measure and Source
Resolution
0.01 mV
Error
Reference Junction Error
Accuracy
±(0.025 % of range (75 mV) + 1 count)
Process Calibrators
Specifications
±0.2 °C
±0.2 °C
11