CALIBRATION
CO
calibration
2
Span setting
Span setting should be done under stable condition of temperature, humidity, and CO
1. Press the calibration key (CAL) once to enter the CO
2. Confirm blinking the decimal point in the digital CO
3. Open the outer door and remove the sample air outlet cap on the inner door.
4. Measure the CO
2
5. Press the calibration key (CAL) for about 5 seconds again to enter the temperature calibration input
mode.
6. The left digit on the digital temperature indicator blinks, and the digital CO
7. Press the calibration key (CAL) once again to enter the CO
8. The left digit on the digital CO
9. Set the present correct CO
then press the enter key (ENT).
10. Replace the sample air outlet cap.
11. The incubator automatically reverts to the display mode.
Note:
This calibration is available when the setting of CO
[Example]
For an internal CO
2
1. Press the calibration key (CAL) once.
2. Press the calibration key (CAL) for about 5 seconds again.
3. The "3" on the digital temperature indicator blinks, and the digital CO
4. Press the calibration key (CAL) once again.
5. The "0" on the digital CO
left digit on the digital CO
6. Set the present correct CO
(
), then press the enter key (ENT).
7. The incubator automatically reverts to the display mode.
Note:
It is important to accurately measure the temperature inside the incubator when calibrating the
temperature. Particularly, the temperature gauge used must have an accuracy of JIS 0.5 Class or better.
The temperature must be measured at several points.
The temperature setting must not change by more than ±1.0
exceeded, an error buzzer will sound, the input data will be ignored, and the incubator will return to the
display mode. If it is necessary to change the temperature by more than 1.0
several stages over a period of time.
density in the chamber through the sample air outlet on the inner door.
density indicator blinks, and the temperature indicator goes out.
2
density with the digit shift key (
2
density of 5.0% (setting) and a measured value of 4.5%.
density indicator blinks, and the digital temperature indicator goes out. The
2
density indicator blinks, and the temperature indicator goes out.
2
density (4.5%) with the digit shift key (
2
calibration pre-conditioning mode.
2
density indicator.
2
calibration input mode.
2
) and numerical value shift key (
density is 2% or more.
2
density indicator goes out.
2
) and numerical value shift key
o
C during calibration. If these values are
32
density.
2
density indicator goes out.
2
o
C, perform calibration in
),