1. Outer door: Sticks to frame with magnetic seal. Door heater is installed in the outer door panel.
The door opening is reversible. Contact Sanyo representative or agent to change the door hinge from
left to right or vice versa.
2. Inner door: Made of tempered glass, however avoid excessive impact on the glass.
3. Leveling foot: Screw type for adjusting the height. Adjust the foot so that the incubator can be
level.
4. Tray: Can be pulled toward you.
5. Tray support: Can be removed by lifting the front side and pulling toward you.
6. Side support: Right and left side supports can be removed for disinfection. See page 27, 28 and
29.
7. Fan cover: Inlet of circulating air. Removable.
8. Duct: Flow path for circulating air. Removable.
9. Fan (inside the rear duct): Made from polypropylene resin. Can be disinfected by an autoclave.
10. Sample air outlet: This also functions as an internal gas outlet. Normally, cover this outlet with
the sample air outlet cap.
11. Sample air outlet cap: Always attach this cap except at the time of using of sample air outlet.
Note: Operation without sample air outlet cap may result low humidity or condensation on the outer door.
12. Connecting port for CO
gas pipe: When an optional component MCO-21GC (automatic CO
2
2
cylinder changeover system) is installed, both A and B are available. If MCO-21GC is not used, only A
is available. Refer to page 14 for gas cylinder connection. Ensure that the gas pressure is set at 0.03
2
MPaG (0.3 kgf/cm
G, 4.3 psiG). Refer to page 39 for automatic CO
cylinder changeover system.
2
13. Door switch:
Detects the door opening/closing and stops the circulating fan, UV lamp* and
electromagnetic valve for CO
when door is open.
2
14. Humidifying pan: Use sterile distilled water to fill the pan. (MCO-18AC(UV) or when an optional
UV system kit MCO-18UVS3 is installed) Install the pan properly so that it can be covered with the pan
cover.
15. Remote alarm terminal: Refer to page 11.
16. Access port: When not in use, cap with the silicon caps on both outside and inside.
17. Power switch: Main switch of the incubator. Also functions as an over-current breaker. The
switch is covered by a switch cover to prevent the accidental push. To turn on or off the switch, remove
the switch cover by loosening the screw.
18. Humidifying pan cover*: Prevents UV light being exposed to the chamber. When filling the pan,
lift the front side and take out the pan. See page 30 for details.
19. UV lamp*: Sanyo UV lamp does not generate ozone. Never look at the UV light directly. For
replacement, contact Sanyo sales representative or agent.
20. Water level sensor for humidifying pan*: Detects the water level in the humidifying pan. See
page 24 for details.
21. Glow starter*: For UV lamp (model; FG-7P)
*MCO-18AC(UV) or when an optional UV system kit MCO-18UVS3 is installed.
9