Setting The Chamber Temperature And Co; Density - Sanyo MCO-18AC UV Instruction Manual

Co2 incubator
Table of Contents

Advertisement

SETTING THE CHAMBER TEMPERATURE AND CO

Table below shows the basic procedure for setting the chamber temperature and CO
upper limit alarm temperature setting is also shown in the table. Perform key operations in the sequence
indicated in the table. The example in the table is based on the assumption that the desired temperature
o
is 37
C and CO
density is 5%. Adjustment of the upper limit regulator should be executed after the
2
chamber temperature reaches the stable condition.
Note:
• The incubator is set at the factory so that the chamber temperature is 37
• When starting the incubator for the first time or after not using it for an extended period of time, allow at
least 8 hours for the chamber temperature, humidity, and CO
desired chamber temperature at 0% CO
Basic operation sequence (Example: Chamber temperature: 33
Description of operation
1
Turn the power switch ON.
2
Press set key.
By pressing numerical value shift
3
key and digit shift key,
set the figure to 37.0.
4
Press enter key.
By pressing digit shift key and
5
numerical value shift key, set the
figure to 05.0.
6
Press enter key.
(Executed after the chamber
temperature reaches the stable
condition)
7
Adjust upper limit regulator so that
the alarm temp. is 1
set temperature.
8
Press enter key.
Note:
● In each set mode, if the change of the setting is not necessary, pressing set key (SET) skips to next set
mode.
● When the CO
density is set to 00.0, the control is OFF regardless of chamber density.
2
● The upper limit temperature set value will change when the regulator is turned even if the incubator is
not in set mode, because the alarm circuit is an independent circuit.
● In each set mode, the indicator returns to the current temperature and CO
automatically when 90 seconds has passed without any key operation.
density. Then change the setting to the desired CO
2
Key operated
----
SET
ENT
ENT
----
o
C higher than
ENT
o
C and CO
sensor to become stable after setting the
2
o
C, CO
2
Indication after operation
The current chamber temperature is displayed
in the digital temperature indicator.
The left digit blinks.
When pressed,
the changeable digit is shifted.
When pressed,
the figure of settable digit changes.
Set temperature is memorized.
Left digit in the digital
CO
density indicator blinks.
2
When pressed, the changeable digit is shifted.
When pressed,
the figure of settable digit changes.
Set CO
density is memorized.
2
In the digital CO
density indicator, HI is displayed.
2
In the digital temperature indicator, upper limit
temp. is displayed.
The upper limit temp. can be changed
by turning upper limit regulator.
This is the end of set mode and the indicators
display current temperature and CO
18

DENSITY

2
density. The
2
density is 0%.
2
density.
2
density: 5%)
density.
2
density display mode
2