Setting The Uv Lamp On Period - Sanyo MCO-18AC UV Instruction Manual

Co2 incubator
Table of Contents

Advertisement

Setting the UV lamp ON period

The clauses below are applicable when the model is MCO-18AC(UV) or an optional UV system kit
MCO-18UVS3 is installed.
Follow the procedure below when changing the setting of the UV lamp ON period.
Basic operation sequence (Example: change of UV lamp ON period from 5 minutes to 3 minutes)
Description of operation
Press calibration key
1
for 5 seconds.
By pressing numerical value shift
2
key and digit shift key,
set the figure to F01.
3 Press enter key.
By pressing numerical value shift
4
key and digit shift key,
set the figure to 003.
5 Press enter key.
 The UV timer can be set from 0 to 30 min. The default setting is for 5 min. When the UV timer is set to
zero, the UV lamp will not light.
 If the outer door is opened while the UV lamp is lit, the lamp will turn OFF. Then, when the door is
closed, the lamp will light for the preset period of time.
 If the UV lamp ON time is set for longer than 5 minutes or if only the outer door is repeated opened and
closed, condensation may occur in the chamber and may affect temperature distribution. It will also
shorten the service life of the UV lamp.
 When replacing the UV lamp, contact a Sanyo sales representative or agent.
Pressing calibration key (CAL) for about 5 seconds leads the calibration mode. In the calibration mode,
the calibration of temperature and CO
performance. Never touch any other keys on the control panel in the event of pressing calibration key
(CAL) accidentally. After about 90 seconds, the incubator returns to chamber temperature display mode
automatically.
Key operated
The left digit in the digital
CAL
temperature indicator is blinked.
When pressed,
the figure of settable digit changes.
When pressed, the changeable digit
is shifted.
The current setting is displayed
ENT
in the digital CO
When pressed, the figure of
settable digit changes.
Set value is memorized and the display return to
ENT
normal display mode.
density is possible. Wrong key operation affects the basic
2
21
Indication after operation
density indicator.
2