Page 2
After the calibration of the board, it is possible to detect with the Primary Open Circuit Test if no transducer is connected. ADDI-DATA GmbH • Dieselstr. 3 • D-77833 Ottersweier • Phone +49(0)72 23 / 94 93-0 • Fax: +49(0)72 23 / 94 93-92 E-Mail: info@addi-data.com...
Page 3
Therefore you must turn the cable. Thus, we recommend you to use the new cable FB3702 – for an easy connection. ADDI-DATA GmbH • Dieselstr. 3 • D-77833 Ottersweier • Phone +49(0)72 23 / 94 93-0 • Fax: +49(0)72 23 / 94 93-92 E-Mail: info@addi-data.com...
Page 5
The content of this manual and the technical product data may be changed without prior notice. ADDI-DATA GmbH reserves the right to make changes to the technical data and the materials included herein. Warranty and liability The user is not permitted to make changes to the product beyond the intended use, or to interfere with the product in any other way.
Page 6
WARNING In case of wrong uses and if the board is not used for the purpose it is intended: ♦ people may be injured, ♦ the board, PC and peripheral may be destroyed, ♦ the environment may be polluted ♦ Protect yourself, the others and the environment! ♦...
Definition of application APCI-3701 DEFINITION OF APPLICATION ........8 Intended use ................8 Limits of use................9 Usage restrictions..............9 USER .................10 Qualification ................. 10 Personal protection............... 10 HANDLING OF THE BOARD.........11 TECHNICAL DATA ............12 Electromagnetic compatibility (EMC) ........12 Physical set-up of the board ..........12 Versions .................
Page 8
APCI-3701 Definition of application SOFTWARE ..............26 Board registration ..............27 7.1.1 Installation of a new board............27 ADDevice Manager..............28 7.1.2 Changing the registration of a board.......... 29 Description of the ADDIREG program .......... 29 Questions and software downloads on the web....32 CONNECTING THE PERIPHERAL........33...
Page 9
Program start ................ 52 11.3 Board information ..............53 11.4 Transducer database ............53 11.4.1 Menu transducer database............54 11.4.2 ADDI-DATA GmbH transducer database........55 11.4.3 User transducer database ............55 11.4.4 New transducer type definition ............55 11.5 Board database ..............58 11.5.1 “Board database” menu..............59 11.5.2 Copy a new transducer type into the board database....59...
Page 10
Fig. 8-6: Digital inputs ................37 Fig. 8-7: Digital outputs................37 Fig. 8-8: Connection to screw terminal panels/boxes ......38 Fig. 9-1: Block diagram of the APCI-3701 ..........39 Fig. 9-2: Acquisition principle of the APCI-3701........39 Fig. 9-3: Half-bridge transducer ............... 43 Fig.
The board APCI-3701 is the interface between an industrial process and a personal computer (PC). The board APCI-3701 must be inserted in a PC with PCI 3.3 V/5V / 32-bit/64-bit slots, which is used as electrical equipment for measurement, control and laboratory use as defined in the norm IEC 61010-1.
APCI-3701 Definition of application Limits of use The APCI-3701 board is not to be used as safety related part for securing emergency stop functions. The emergency stop functions are to be secured separately. This securing must not be influenced by the board or the PC.
User APCI-3701 USER Qualification Only persons trained in electronics are entitled to perform the following works: installation use, maintenance. Personal protection Consider the country-specific regulations about: the prevention of accidents electrical and mechanical installations radio interference suppression.
Installation in: ........... 32/64-bit PCI slot 5 V or 3.3 V Connection to the peripheral: ....50-pin SUB-D male connector Accessories for the standard version APCI-3701 Cables: Standard cable: ......... ST3701, 2 m; max. length: 10 m Ribbon cable: ..........FB3702 (connection of the dig. I/O) Connection boxes/Screw terminal panel: for LVDT transducers: ......
APCI-3701 Technical data Versions The board APCI-3701 is available in the following versions: Version Number of connected Number of connected Standard transducers digital I/O APCI-3701-16 8 inputs and 8 outputs APCI-3701-8 8 inputs and 8 outputs Please observe: If you use LVDT transducers ≥±1 mm, please use the version APCI-3701-LVDT+.
Technical data APCI-3701 Energy requirements: - Operating voltage of the PC: ....5 V ± 5% - Current consumption (without load): ..typ. see table ± 10% APCI-3701-8 APCI-3701-16 0.48 A 0.58 A + 5 V from PC WARNING ! The cooling components and the DC/DC connector can warm up during the board operation.
APCI-3701 Technical data 4.4.2 Analog inputs Number: ............ 8 to 16 Input type: ..........single-ended Coupling: ........... AC Edge frequency of the high-pass filter (-3 dB):......500 Hz Edge frequency of the low pass filter (-3 dB): ....... 32 kHz Maximum input voltage: ......
Technical data APCI-3701 Technical data of the accessories 4.5.1 Connection boxes and cables of the standard versions Version Number of connected transducers Connection cable PX3701HB-16 16 half-bridge transducers ST3701-16 PX3701HB-8 8 half-bridge transducers ST3701-8 PX3701LVDT-16 16 LVDT transducers ST3701-16 PX3701LVDT-8...
If you want to install another transducer that is not included in this library, contact us. We will adapt the board so that you can run your transducer with the required parameters and data through the ADDI-DATA APCI-3701. ADDI-DATA sales department: + 49 7223 94 93 -120 Load Transducer...
APCI-3701 Installation of the board INSTALLATION OF THE BOARD IMPORTANT! Do observe the safety precautions (yellow leaflet)! Opening the PC ♦ Switch off your PC and all the units connected to the PC ♦ Pull the PC mains plug from the socket.
Installation of the board APCI-3701 Take the board out of its protective pack. ♦ Plugging the board into the slot into the chosen slot. Fig. 5-2: Inserting the board...
APCI-3701 Installation of the board ♦ Fasten the board to the rear of the PC housing with the screw which was fixed on the back cover. Fig. 5-3: Fastening the board at the back cover ♦ Tighten all the loosen screws.
On the next page you will find a overview of the cables (including the combination possibilities with the connection boxes and the different versions of the board APCI-3701). In addition, in the following sections you will find more detailed information...
Page 26
APCI-3701 Accessories Boards Connection boxes Standard connection boxes Cable type Order name APCI- APCI- APCI- APCI- APCI- PX3701- PX3701- PX3701- PX3701- 3701-1 3701-8 3701-16 3701-8-K 3701-16- HB-8 HB-8 LVDT-8 LVDT-16 Length Standard cable (chapter Fehler! ST3701 ▲ ■ ▲ ■...
The standard cables are considered for the use of standard transducers, e.g. Tesa, Schaevitz, etc. in common industrial environment. 6.1.2 1-channel-cables: ST3701-1-HB und ST3701-1-LVDT The standard delivery of the APCI-3701-1 always contains one of the both 1-channel-connection cables (whereas, additionally, both cable types can be ordered separately).
Connection examples Connection: 1-channel version with a transducer For the connection of the APCI-3701-1, please use either the cable ST3701-1-HB or the cable ST3701-1-LVDT with round connector. Connect this cable with the APCI-3701 and with the transducer. You do not require any additional terminal panels.
ADDI-DATA software is based on the principle of a virtual board: it transposes the different functions (e.g. digital inputs, analogy outputs, timer, ...) of all inserted ADDI-DATA boards as the functions of a single (virtual) board. The virtual board features a pool of functions, the functionality of which can be called up without calling a specific board.
APCI-3701 Software Board registration When starting the set-up of ADDIREG, the APCI-3701 is automatically recognised and registered. 7.1.1 Installation of a new board If a new board is recognised, the following window is displayed: Fig. 7-1: New inserted board (example)
Software APCI-3701 ADDevice Manager Fig. 7-2: ADDevice Manager (example) The following parameters are displayed for every inserted board: First column: Board name Board index: Number allocated to the board when it is registered in ADDIREG. Slot number IRQ line Different addresses which are automatically allocated to the board by the BIOS.
APCI-3701 Software By clicking twice within a column, the connection principle and the technical data of the resource are displayed. This function is only possible if a question mark appears with the cursor. You can export the set configuration as a text file. Click on "file" and save the configuration as a .txt file with "Export information to file...".
Page 33
Selected base address of the board. For PCI boards the base address is allocated through BIOS. Access: Selection of the access mode for the ADDI-DATA digital boards. Access in 8-bit or 16-bit or 32-bit mode. PCI bus/device/(slot): Number of the used PCI bus, slot, and device. If the board is no PCI board, the message "NO"...
Page 34
APCI-3701 Software Buttons: Edit: Selection of the highlighted board with the different parameters set in the text boxes. Set: Sets the parameterised board configuration. The configuration should be set before you save it. Cancel: Reactivates the former parameters of the saved configuration.
Software APCI-3701 Questions and software downloads on the web Do not hesitate to e-mail us your questions. per e-mail: info@addi-data.de hotline@addi-data.de Free downloads of standard software You can download the latest version of the software for the board APCI-3701. http://www.addi-data.com...
APCI-3701 Connecting the peripheral CONNECTING THE PERIPHERAL Connector pin assignment 8.1.1 50-pin SUB-D male front connector ST1 Fig. 8-1: Pin assignment: APCI-3701-16 Signal Meaning OSC+ (Red pins) or phase-shifted supply signal of the inductive transducers OSC- (Blue pins) BACK+ (green pins) or...
Connecting the peripheral APCI-3701 Fig. 8-2: Pin assignment: APCI-3701-8 NC: not connected 8.1.2 40-pin male connector ST3 Fig. 8-3: Pin assignment: 40-pin connector to 37-pin SUB-D bracket IMPORTANT! Plug the FB3702 on the connector with the red cable lead on...
APCI-3701 Connecting the peripheral Connection of inductive transducers Two connection boxes with 8 (PX371-8) or 16 (PX371-16) transducer female connectors and the belonging connecting cable are available for connecting the inductive displacement transducers to the board. The connection with the board is made via a 50-pin SUB-D male connector.
Connecting the peripheral APCI-3701 8.2.3 Connection of Mahr transducers The pin assignment for Mahr transducers of the product range P2xxx-M (e.g.: P2004-M) differs from the pin assignment for standard half-bridge transducers (e.g. GT21 of Tesa). Should you want to connect a Mahr transducer to the Box PX3701-HB-x with the Mahr assignment (Designation M), an adapter between the transducer and the flange socket of the connection box is necessary.
Connecting the peripheral APCI-3701 Connection to screw terminal panels/boxes Fig. 8-8: Connection to screw terminal panels/boxes IMPORTANT! Plug the FB3702 on the connector with the red cable lead on the side of the pin 1. The cable ST3701 is 2 m long in the standard delivery; The max. permissible...
♦ Install the board in the PC so that cooling air can flow easily between the different inserted boards. ♦ Before removing the board from the PC, wait until it sufficiently cooled down. Block diagram Fig. 9-1: Block diagram of the APCI-3701 Acquisition principle Fig. 9-2: Acquisition principle of the APCI-3701...
Functions of the board APCI-3701 The board APCI-3701 has up to 16 analog inputs. It is typically designed for acquiring measurement signals of inductive displacement transducers. Therefore the board provides all signals required for supplying the inductive transducers. The primary circuit of the sensor is excited by means of a sinus generator.
APCI-3701 Functions of the board Primary circuit 9.3.1 Sinus generator The sinus generator generates a sinus signal for the excitation of the primary circuit. The frequency and the amplitude are set through software according to the connected sensor type. The frequency can be set with 14 predefined values in a range from 2 to 20 kHz (See Table 4-1).
Functions of the board APCI-3701 Secondary circuit The incoming measuring signals are driven through 2 separate 8:1 multiplexers. Each input level is then amplified by a programmable amplifier. The measuring signal is driven through analogue filters (pass band filters) with edge frequencies of 500 Hz and 32 kHz and acquired simultaneously with the feedback signal (return line signal) in the A/D converter.
APCI-3701 Functions of the board Transducer principle The inductive transducer is basically used for precise measuring travels. It is a travel/voltage sensor, the output voltage of which changes linearly along with the moving ferromagnetic core. The ferromagnetic core moves according to a straight line in a transformator, which consists of a central primary coil and 2 external secondary coils (cylindrical windings).
Functions of the board APCI-3701 9.5.2 LVDT transducer (Series-opposed) The series-opposed transducer features 3 inductive coils, one primary coil and 2 secondary coils. These coils are positioned concentrically around the mobile core and form two symmetrical transformers with respect to the electrical zero point of the transducer.
APCI-3701 Functions of the board Possible acquisitions The board APCI-3701 generates the sinusoidal supply voltage and evaluates the measuring signal. Each channel can be independently acquired (single mode; See Fig. 9-5) The board can acquire the channels in sequences: n sequences can be carried out; 1 ≤ n ≤ 65.535 (See Fig. 9-7) n sequences can be divided in k sequences.
Functions of the board APCI-3701 Fig. 9-5: Possible acquisitions: Single start enable Fig. 9-6: Possible acquisitions: single software/hardware trigger Fig. 9-7: Possible acquisitions: sequence start enable Fig. 9-8: Possible acquisitions: m/n sequences with hardware/software trigger...
APCI-3701 Functions of the board Diagnostic The board disposes of a diagnostic function which under certain circumstances can detect a short circuit or an open line on the primary circuit as well as on the secondary circuit. The short-circuit detection on the primary circuit is activated continuously. The other diagnostic functions are activated by software functions.
Standard software APCI-3701 STANDARD SOFTWARE 10.1 Software functions ADDIPACK supports the following functions for the APCI-3701. Table 10-1: Supported software functions Functionality Function name i_ADDIDATA_OpenWin32Driver Common Functions i_ADDIDATA_GetCurrentDriverHandle i_ADDIDATA_GetDriverVersion b_ADDIDATA_CloseWin32Driver Interrupt b_ADDIDATA_SetFunctionalityIntRoutineWin32 b_ADDIDATA_TestInterrupt b_ADDIDATA_ResetFunctionalityIntRoutine Error i_ADDIDATA_GetLastError i_ADDIDATA_GetLastErrorAndSource b_ADDIDATA_EnableErrorMessage b_ADDIDATA_DisableErrorMessage b_ADDIDATA_FormatErrorMessage...
APCI-3701 Standard software 10.2 Software samples Table 10-2: Supported software samples Functionality Sample Description number SAMPLE00 Show information of a transducer channel Transducer SAMPLE01 Read 1 transducer channel without interrupt. SAMPLE02 Read 1 transducer channel with interrupt. SAMPLE03 Several transducer channels without interrupt.
SET3701: CALIBRATION AND UPDATE TOOL 11.1 Introduction This program allows you to calibrate the APCI-3701 with any transducer type. ADDI-DATA GmbH selected the most common used transducer types, but you may define any other type. This software consists of two parts: 1.Transducer database management...
Board base address Board interrupt Board serial number Revisionsnummer der Firmware If you have several APCI-3701 boards in you PC: ♦ Double-click on the board icon An animated menu opens. ♦ Double-click on the required board to select it 11.4 Transducer database ♦...
User transducer database database Selected transducer characteristics The APCI-3701 can be connected to different transducers. All transducer information will be saved in two databases: - Addi-Data GmbH transducer database - User transducer database Each database contains the following transducer information:...
“Addi-Data GmbH transducers database”, you shall define a new transducer type. The new types will be saved in database file. The file name for the standard board APCI-3701 is „User_SV.DB“ and for the Knäbel version of the APCI-3701 „User_KV.DB“.
SET3701: Calibration and update tool APCI-3701 A new window appears. You must complete the following fields: - Name - Type (LVDT orr HB) - Sensitivity (mV/v/mm) - Veff (Vrms) - Frequency (Hz) - Range (mm) - Load impedance (KOhm) The characteristics of the transducer shall be given by the transducer manufacturer.
APCI-3701 SET3701: Calibration and update tool For modifying the transducer information: ♦ Double-click on the required type or use the menu “Change type”. ♦ If all information is correct, click on “Save database” to save the new type. Fig. 11-5: Save a new type A new window appears: The version above the „User...
All available transducer the board types Calibration information Characteristics of the selected transducer You can load max.14 transducer types on an APCI-3701. “Transducers database” listes all available transducers. This list contains the “Addi-Data Gmbh transducer database” (ADDI-DATA GmbH and the “User...
APCI-3701 SET3701: Calibration and update tool transducer database”. (See chapter 11.4). “Board database” lists all transducer types that are loaded on the APCI-3701. 11.5.1 “Board database” menu Copies the selected Abb. 11-1: “Board database” menu transducer type from the transducer database to the board database.
SET3701: Calibration and update tool APCI-3701 IMPORTANT! You can copy only 14 transducer types to the „Board database“. Transducer types that are not used, you can delete with the command „Del type“ to leave space for the other ones. 11.5.4 Copy a transducer type through drag & drop ♦...
SET3701: Calibration and update tool 11.5.5 Save the board database After having loaded all required transducer types, select “Save database” to load this list on the selected APCI-3701. 11.6 Transducer calibration ♦ Double-click on the main menu icon to activate the window “Transducer calibration”...
Calibration steps For each used transducer type you have to calibrate the board. The “Transducer type selection” lists all transducer types that are loaded on the APCI-3701. “Calibration position” indicates the position that is used for calibration. This position ranges from 0 to the maximum positive transducer position. Per default the software selects the middle of this range.
For the calibrationn you need a PX 3701-Box and the cable ST3701. Connect the box to the APCI-3701 via the cable. After having selected the transducer type that you want to calibrate, the calibration position and the channels of group 1 and group 2 to which the transducers are connected, click on “Start calibration“.
Page 67
APCI-3701 If this step takes longer or if an error message appears, this means that you have not connected the PX 3701 to the APCI-3701. In this case, stop the calibration with “Break calibration”, connect the PX 3701 and start again with the calibration.
Page 68
APCI-3701 SET3701: Calibration and update tool ♦ Click on “OK” ♦ Set the transducer to the mechanical selected calibration position and click on “Next” The calibration is completed if you have only 8 channels. If you have 16 channels, the same steps are started for the second channel group.
When the window first appears, the last calibrated transducer type is selected automatically as default and all available channels are selected. You can change the transducer type. All transducer types that are loaded on the APCI-3701 can be selected.
APCI-3701 SET3701: Calibration and update tool 11.7.1 Transducer monitoring menu Fig. 11-10: Monitoring menu Starts the monitoring Stops the monitoring Selects all channels Unclicks all channels Transducer monitoring help 11.7.2 Acquisition You can select one or more channels for the monitoring. For each channel you can select any available transducer type.
SET3701: Calibration and update tool APCI-3701 11.8 Firmware update ♦ Double-click on the main menu icon to call the window “Firmware update” The animated window appears. ♦ Double-click on “Firmware update”. The animated menu disappears and the window “Firmware update” is activated.
Fig. 11-11: Window „Firmware update“ Transducer type selection A firmware update is required if you have a standard APCI-3701 with a firmware revision lower than 2.6. If you start the calibration and the board firmware is from a lower version than the required version, the program automatically runs an update.
Glossary APCI-3701 GLOSSARY Table 12-1: Glossary Term Description A/D converter = ADC An electronic device that produces a digital output directly proportional to an analog signal output. Acquisition The process by which data is gathered by the computer for analysis or storage.
Page 74
APCI-3701 Glossary Typically the half-bridge transducer forms half of a Wheatstone Bridge circuit which enables change from null to be readily determined. Impedance The reciprocal of admittance. Admittance is the complex ratio of the voltage across divided by the current flowing through a device, circuit element, or network.
Page 75
Glossary APCI-3701 Optical isolation The technique of using an optoelectric transmitter and receiver to transfer data without electrical continuity, to eliminate high- potential differences and transients. Opto-coupler A device containing light-emitting and light-sensitive components used to couple isolated circuits Output voltage...
Accesories list 28 Interrupt 54 Accessories 13, 27 Acquisition 52 Acquisition principle 46 ADDevice Manager 35 Limit values of the version APCI-3701-K 18 ADDIDRIVER 33 Limits of use 10 ADDIPACK 33 LVDT transducer ADDIREG registration program 36 Function description 51...
Page 77
Index APCI-3701 Transducer list 22 Qualification 11 Transducer principle 50 Versions 15 Update 39 Virtual board 33 Usage restrictions 10 User Personal protection 11...