29.6.15 Query lower tolerance unit
Command: ODH CR LF
˽
Response:
DH
MASA CR LF
1
2
D
H
Weight:
- 9 characters right aligned
Unit:
- 3 characters left aligned
29.6.16 Query upper tolerance unit
Command: OUH CR LF
˽
Response:
UH
MASA CR LF
1
2
U
H
Weight:
- 9 characters right aligned
Unit:
- 3 characters left aligned
29.6.17 Enter reference single weight
˽
Command: SM
XXXXX CR LF
Possible replies:
˽
SM
OK CR LF
˽
SM
I CR LF
ES CR LF
29.6.18 Enter target weight (dosing)
˽
Command: TV
XXXXX CR LF
Possible replies:
˽
TV
OK CR LF
˽
TV
I CR LF
ES CR LF
AEJ-C/AES-C/PLJ-C-BA-e-1621
Command executed
3
4 - 12
13
˽
˽
Tare
Command executed
3
4 - 12
13
˽
˽
Tare
Command executed
Command valid but cannot be executed at the moment (e. g.
incorrect operating mode)
Syntax error (incorrect weight format)
Command executed
Command valid but cannot be executed at the moment (e. g.
incorrect operating mode)
Syntax error (incorrect weight format)
14 - 16
17
˽
Unit
14 - 16
17
˽
Unit
18
19
CR
LF
18
19
CR
LF
165