29.6.8
Send weighing value immediately in current weighing unit
Command: SUI CR LF
Possible replies:
˽
SUI
I CR LF
1
2
3
4
Stability
S
U
I
sign
Example:
S U I
CR LF
˽
˽˽˽
-
S U I ?
5 8 . 2 3 7
29.6.9 Start continuous output in standard weighing unit
Command: C1 CR LF
Possible replies:
˽
C1
I CR LF
˽
C1
A CR LF
1
2
3
4
Stability
˽
S
I
sign
29.6.10 Stop continuous output in standard weighing unit
Command: C1 CR LF
Possible replies:
˽
C0
I CR LF
˽
C0
A CR LF
AEJ-C/AES-C/PLJ-C-BA-e-1621
Command valid but cannot be executed at the moment
5
6
˽
Signs
Sent command, see chap. 29.4
˽
˽
Command executed.
k g
CR LF
in current weighing unit
Command valid but cannot be executed at the moment
Command accepted, command executing
Weighing values are shown in default weighing unit
5
6
˽
Signs
Command valid but cannot be executed at the moment
Command accepted, command executing
7-15
16
17-19
˽
Weight
Unit
Weighing value is shown
7-15
16
17-19
˽
Weight
Unit
20
21
CR
LF
20
21
CR
LF
163