29.6 Detailed information to communication logs

29.6.1

Zeroing

Command: Z CR LF
Possible replies:
˽
Command accepted, command executing
Z
A CR LF
˽
Command execution completed
Z
D CR LF
˽
Command accepted, command executing
Z
A CR LF
˽
Z
^ CR LF
Command valid but zero reset range was exceeded
Command accepted, command executing
˽
Z
A CR LF
Time limit for stabilisation of weighing balance display was
˽
Z
E CR LF
exceeded
˽
Command valid but cannot be executed at the moment
Z
CR LF
29.6.2

Taring

Command: T CR LF
Possible replies:
˽
Command accepted, command executing
T
A CR LF
˽
Command execution completed
T
D CR LF
˽
T
A CR LF
Command accepted, command executing
˽
Command valid but tare range was exceeded
T
^ CR LF
Command accepted, command executing
˽
T
A CR LF
Time limit for stabilisation of weighing balance display was
˽
T
E CR LF
exceeded
˽
Command valid but cannot be executed at the moment
T
I CR LF
29.6.3

Query tare value

Command: OT CR LF
Response:
1
2
O
T
Tare value: - 9 characters right aligned
AEJ-C/AES-C/PLJ-C-BA-e-1621
3
4 - 12
13
˽
˽
Tare
14 - 16
17
˽
Unit
18
19
CR
LF
159