P6.9
DATE FORMAT
(on logs)
P6.10
TIME FORMAT
(on logs)
P6.11
GLP AUTO TEST
Additional log output by pressing
You can cancel the test by pressing
Press back to weighing mode
56
Options:
YYYY.MM.DD / YYYY.DD.MM / DD.MM.YYYY / MM. DD.YYYY
Options: 12H / 24H
The selected format AM or PM will be printed on the log next to
the time.
Repeatability test activated by placing the internal weight
Sequence of operations:
To start the test, confirm the function <P6.9
AUTOTEST GLP> by pressing
The internal weight will be applied twice, followed by
10 repeat measurements.
Adjustment
Calculation of standard deviation
Automatic log printout
Printout example KERN YKB-01N
-------------------
GLP Autotest Rep.
Scale Type
Balance ID
User
Soft. Ver.
Date
Time
-------------------------------------------------------------------
Measur. No.
Division
Int. Weight mass
Filter
Result cong.
-------------------------------------------------------------------
Max. Dev.
Repeatab.
Signature
..................................................................
.
-------------------
PL
150854
Admin
r3.2.6
05.01.2016
11:34:17
10
0.01 g
765.35
Means
Fast+rel.
0.01 g
0.003 g
.
.
several times.
AEJ-C/AES-C/PLJ-C-BA-e-1621