Probe Specifications
Type
Calibration
Process Temperature
Storage Temperature
Operating Pressure
Flow Range
Moisture Image Series
Probe/Analyzer
Separation
Moisture Image Series
Probe/Analyzer Cable
Built-in Temperature
Sensor (optional)
Specifications
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Moisture Image Series Probe
Aluminum oxide moisture sensor probe and electronics module.
Each sensor is individually computer calibrated against known
moisture concentrations, traceable to National Institute of Standards
and Technology (NIST) or National Physical Lab, U.K. (NPL) as
approved by Irish Laboratory Accreditation Board (ILAB).
-110°C to 70°C (-166°F to 158°F) dew/frost point temperature
70°C (158°F) maximum
5 microns of Hg to 5000 PSIG
Gases:
Static to 10,000 cm/sec linear velocity at 1 atm
Liquids:
Static to 10 cm/sec linear velocity at 1 g/cc
Response Time:
<5 seconds for 63% step change in moisture content in either wet up
or dry down cycle.
915 meters (3000 feet) with AWG 22 twisted-pair cable.
Unshielded, twisted pair, maximum loop resistance of 100 ohms
Type:
Thermistor network
Operating Range:
-30°C to 70°C (-22°F to 158°F)
Accuracy:
±0.5°C overall
Response Time:
1 second in well-stirred oil
10 seconds in still air for a 63% step change in increasing or
decreasing temperatures.
April 2004
3-7