Weatherproof Delta F Oxygen Cells; Weatherproof Delta F Oxygen Cell Connections; Weatherproof Delta F Oxygen Cell - GE Moisture Image Series 1 Startup Manual

Hygrometer
Hide thumbs Also See for Moisture Image Series 1:
Table of Contents

Advertisement

Weatherproof Delta F
Oxygen Cells
Installation
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Figure 1-19 below shows the weatherproof oxygen cell. The oxygen
cell has a set of sensing and secondary electrodes that are wired to a
terminal strip in the weatherproof enclosure. Connect the
weatherproof oxygen cell using a four-wire shielded cable with a
mating bayonet-type connector. Fasten the bayonet-type connector to
the mating connector on the bottom of the weatherproof enclosure.
Connect the other end of the cable to the
the back of the electronics unit as shown in Table 1-7 below.
IMPORTANT: To maintain good contact at each terminal block and
to avoid damaging the pins on the connector, pull the
connector straight off (not at an angle), make cable
connections while the connector is away from the
unit, and push the connector straight on (not at an
angle) when the wiring is complete.
Table 1-7: Weatherproof Delta F Oxygen Cell Connections
To Delta F Enclosure
Connect:
Terminal Block
red wire (+)
green wire (–)
white wire (+)
black wire (–)
shield
Enclosure
Terminal
Block
Figure 1-19: Weatherproof Delta F Oxygen Cell
OXYGEN
terminal block on
To Series 1 OXYGEN
Terminal Block
pin 1
pin 3
pin 4
pin 5
pin 2
June 2004
pin 1
pin 2
pin 3
pin 4
pin 5
1-31