April 2004
Moisture Image Series
Probes (cont.)
1-26
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Once the probe is assembled, connect the cable to the terminal block
labeled
on the back panel of the electronics unit (refer to
MIS PRB
Figure 1-15 and Table 1-4 below).
IMPORTANT: DO NOT connect the cable to the Moisture Image
Series Probe until you make the proper connections
to the back of the meter.
IMPORTANT: To maintain good contact at each terminal block and
to avoid damaging the pins on the connector, pull the
connector straight off (not at an angle), make cable
connections while the connector is away from the
unit, and push the connector straight on (not at an
angle) when the wiring is complete.
You can connect the Moisture Image Series Probe to any channel.
However, if you are also using other sensors, such as the M and/or TF
Series probes, be sure to connect the Moisture Image Series Probe to
an open channel.
IMPORTANT: Check the Calibration Data Sheets (of all the
sensors) to determine which channels already have
probe assignments.
Table 1-4: Moisture Image Series Probe Connections
Connect:
positive wire (white)
return wire (black)
MIS PRB
Terminal Block
Figure 1-15: MIS PRB Terminal Block Location
To MIS PRB Term. Block:
pin #1
pin #2
Installation