Entering Moisture
Reference Values (cont.)
2-18
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
HIGH
1. Move the pointer to
[YES].
2. Enter the high reference value for that channel and press [YES].
3. The pointer automatically moves to
Press [YES].
4. Enter the low reference value for that channel, and press [YES].
5. Repeat steps 1 through 4 to enter the high and low reference
values for the remaining channels.
Note: You may also enter the Auto-Cal Interval while in the
Reference Menu. Simply press the INTERVAL menu key and
enter the desired interval. See Chapter 3 of the Programming
Manual for more information.
To enter reference values for oxygen cell inputs, refer to the
following section. To exit, press the DONE menu key until Main
Menu appears on the message line. Then refer to Displaying
Measurements on page 2-21. When you exit, the Series 1
automatically calibrates each channel.
for the desired channel, and press
LOW
for the same channel.