GE Moisture Image Series 1 Startup Manual page 73

Hygrometer
Hide thumbs Also See for Moisture Image Series 1:
Table of Contents

Advertisement

April 2004
Entering Delta F Oxygen
Cell Reference Values
(cont.)
2-20
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
1. Move the pointer to the
desired channel and press [YES].
2. Enter the zero value for the high reference for that channel and
press [YES]. The pointer automatically proceeds to the
for the high reference for the same channel. Press [YES] again.
3. Enter the span value for the high reference for that channel and
press [YES].
4. Move the pointer to the
desired channel and press [YES].
5. Enter the zero value for the low reference for that channel and
press [YES].
6. Move the pointer to the
desired channel and press [YES].
7. Enter the span value for the low reference for that channel and
press [YES].
Repeat steps 1 through 7 to enter the high and low reference values
for the remaining channels.
To enter reference values for moisture inputs, refer to A. Entering
Moisture Reference Values, page 2-17. (To enter reference values for
other inputs, refer to Chapter 2 of the Programming Manual.) To exit,
press the DONE menu key until
line. Then refer to Displaying Measurements on page 2-21. When you
exit, the Series 1 automatically calibrates each channel.
Zero
line for the high reference for the
Zero
line for the low reference for the
Span
line for the low reference for the
Main Menu
appears on the message
Span
line
Initial Setup