3.8.2
Methode "MLAT"
Run sequences for the 838 Advanced Sample Processor
838 Advanced S.Proc.
parameters
method
number of samples:
>start sequence
1 CTL:Rm:
2 MOVE
1
:
3 CTL:Rm:
*************1
4 CTL:Rm:
*************0
>sample sequence
1 SCN:Rm
:
2 DOS: 1.* :
3 DOS: 2.* :
4 CTL:Rm:
*************1
5 CTL:Rm:
*************0
6 DOS: *.* :
7 SCN:Rm
:
8 MOVE
1
:
9 LIFT: 1
:
10 PERISTALT: 300 s
11 CTL:Rm:
*************1
12 CTL:Rm:
*************0
13 MOVE
1
:
14 LIFT: 1
:
15 PERISTALT:
16 SCN:Rm
:
17 PERISTALT: 300 s
18 CTL:Rm:
*************1
19 CTL:Rm:
*************0
>final sequence
>changer settings
>dosing unit def.
>timeout settings
>manual stop
Metrohm 838 Advanced Sample Processor, Appendix
MLAT = Modified Linear Approximation Technique
Application
Brightener analysis
System setup
See 8.838.1303 Installation Instructions, Section 3.6 or 8.797.1013
Hardware Manual of the 797 VA Computrace, Section 2.11.4.
00030 5.838.0010
MLAT
rack
INIT
sample
*****1**
30 mL
1 mL
fill mL
*****1**
sample
work mm
10
+28
work mm
5 s
10
*****1**
10
Note: The command lines 1 to 6 are only necessary, if dosing drives
are connected to the Sample Processor for adding auxiliary solutions.
In the 797 VA Computrace software the parameter "Dose auxiliary so-
lution via sample processor" in the configuration dialog "Settings /
General settings / Automation" must be activated.
If adding of auxiliary solutions is operated by the 797 VA Com-
putrace, line 1 to 6 have to be deleted.
- report header with instrument ID and program version
- method name
- number of samples (e. g. entire rack)
- initialize all Remote lines
- approach first sample
- remote start of the 797 VA Computrace
- wait for advance signal of the 797 VA Computrace
- add VMS
- add suppressor solution
- set Remote line: Intercept solution ready
- fill dosing units
- wait for advance signal of the 797 VA Computrace
- approach sample
- run lift and needle to work position
- transfer sample with peristaltic pump
- set Remote line: sample transfer ready
- approach rinsing vessel
- run needle to work position
- aspirate rinsing solution
- wait for advance signal of the 797 VA Computrace
- rinse needle and transfer tubing
- set Remote line: needle cleaning ready
----- miscellaneous settings -----
3.8 Standard methods
83