2.2 Configuration
2.2.5
RS232 settings
The correct settings of the interface parameters of the serial RS232 interface is essential for the
perfect functioning of data transmission to and from the Sample Processor. This includes the
control of the instrument by using a PC software such as «IC Net».
configuration
>RS232 settings
>RS232 settings
baud rate:
Use <QUIT>
for next higher
level
>RS232 settings
data bit:
>RS232 settings
stop bit:
>RS232 settings
parity:
>RS232 settings
handshake:
>RS232 settings
character set:
For data transmission using personal computers you should select
>RS232 settings
RS control:
If Remote control is switched off then no data will be received.
24
9600
300,600,1200,
2400,4800,9600, 19200
7,8
1,2
none
even, odd,
none
HWs
HWs,SWchar,SWline,
none
IBM
IBM,HP,Epson,
Seiko,Citizen
ON,OFF
Submenu for serial interface settings
Open the submenu with <ENTER>
Transmission rate
in baud
Number of data bits
8
Number of stop bits
1
Parity selection
Handshake selection
Character set for printer and PC
(printer emulation)
Switch data reception on/off
ON
Metrohm 838 Advanced Sample Processor, Operation
.