2.7 Manual operation
LIFT
9
2.7.3

Setting the sample position

SAMPLE
7
2.7.4

Pump control

PUMP
4
2.7.5

Switching the stirrers

STIR
5
50
manual operation
LIFT: 1
:
work,rinse,shift,
special,rest,
With the LIFT command, the lift can be run to a given position.
In addition to selecting an exact position in mm (0 – 125 mm)
the <SELECT> key can select a predefined position (work po-
sition, rinse position, shift position, special position, rest position
= 0 mm). The lowest possible lift position can be set in the con-
figuration menu, see section 2.2.2.
manual operation
SAMPLE:
=
The <SAMPLE> command serves to set the current sample posi-
tion. It defines the first sample tube for the subsequent sample se-
ries.
manual operation
PERISTALT:
on s
on,off
-15...-1,1...15
The PUMP command is used to switch on/off the peristaltic pump at
the 838 Advanced Sample Processor. The PERISTALT command
switches on or off the peristaltic pump, according to the current
state. Pump rate and rotation direction of the pump drive can be set
in 15 steps. A negative value of the pump rate defines a counter-
clockwise rotation. A positive value of the pump rate defines a
clockwise rotation.
manual operation
STIR: T1
:
T1, T2, T*,
MSB1, MSB2,
MSB 3; MSB*,*
on, off,
1...9999 s
Position the lift
rest mm
0...235 mm
Define current sample
1
position
1...999
Switch peristaltic pump
3
on/off
Stirrer control
1 s
Metrohm 838 Advanced Sample Processor, Operation