5 Operation

5.2 Start up the instrument

5.2
Start up the instrument
Precautions
Turn on power and run
diagnostic program
Step
1
2
3
4
66
WARNING
Check that the high voltage is turned off before opening the safety
lid. The HV ON indicator on the front panel should not light.
WARNING
In a situation where there is a risk of injury, turn off the instrument
by switching the mains power switch (located on the rear panel)
to position 0. Malfunctions must be rectified before Ettan IPGphor 3
Isoelectric Focusing System is restarted.
Action
Turn on the mains Power switch located on the rear panel.
This activates a self diagnostic program that runs for approximately 10
seconds. A series of screens indicate progress. If any component fails, the
diagnostic program will stop and a message will indicate the source of the
fault.
Note the fault, if any.
Press START to skip to the next test.
Call GE service if any failure is detected,
Ettan IPGphor 3 Operating Instructions 28-9635-51 AC