GE Ettan IPGphor 3 Operating Instructions Manual page 78

Hide thumbs Also See for Ettan IPGphor 3:
Table of Contents

Advertisement

5 Operation
5.3 Perform a run from the control panel
5.3.3 Set protocol parameters
Step
7
8
9
78
Action
Adjust the current limit for all isoelectric focusing steps by using the Left
and Right arrow keys to position the cursor under one of the three active
μA digits, and use the Up and Down arrow keys to select the value within
the range of 0 to 200 μA per strip.
The original factory programmed value is 50 μA per IPG strip.
P r o t # 1 F i l e 1
R e h y d r a t e 1 0 : 0 0 H r @ 2 0 ˚ C
I E F @ 2 0 ˚ C 5 0 µ A / S t r i p
1 S t r i p s 0 S t e p s
Note:
The total deliverable ampeerage is 1500 μA.
Use the arrow keys to adjust the numbers of strips to run.
P r o t # 1 F i l e 1
R e h y d r a t e 1 0 : 0 0 H r @ 2 0 ˚ C
I E F @ 2 0 ˚ C 5 0 µ A / S t r i p
1 S t r i p s 0 S t e p s
Press EDIT on the control panel to move to next screen.
Ettan IPGphor 3 Operating Instructions 28-9635-51 AC