5.3.1

Ettan IPGphor 3 control panel functionality

Introduction
IEF runs are programmed from theEttan IPGphor 3 Isoelectric Focusing System control
panel with up to ten user-defined IEF protocols, each with up to 9 steps per protocol.
Programmable functions including: rehydration time, platform temperature, current
limit, voltage limit for each step, voltage gradient or step and step duration.
Control unit keys
The following keys on the control panel are used to program Ettan IPGphor 3.
COMMAND
keys
EDIT
Left arrow
Right arrow
Ettan IPGphor 3 Operating Instructions 28-9635-51 AC

5.3.1 Ettan IPGphor 3 control panel functionality

Function
Pressing the EDIT key in the edit mode toggles between the proro-
col identification screen with general settings for all steps and the
settings in the three first protocol steps.
The EDIT key switches the display to edit mode, in which protocol
steps can be programmed. (Refer to Section 5.3.3 Set protocol pa-
rameters, on page 75.
The EDIT key also allows the estimated run time remaining to be
viewed while the unit is operating.
The Left arrow key moves the cursor to the left. Moving the cursor
off the screen to the left repositions the cursor at the left side of
the next row. From the bottom row the cursor is repositioned on
the left side of the top row.
The Right arrow key moves the cursor to the right. Moving the
cursor off the screen to the right repositions the cursor at the left
of the next row. From the bottom row the cursor is repositioned
on the left side of the top row.
5.3 Perform a run from the control panel
5 Operation