Keysight N777 C Series Programming Manual page 70

Tunable laser source
Table of Contents

Advertisement

4
Measurement Operations & Settings
Response:
The number of cycles as a signed integer value.
Example:
:sour0:wav:swe:cycl? -> +3
Affects:
All N777xC tunable laser sources
Command:
:SOURce0:WAVelength:SWEep:DWELl
Syntax:
:SOURce0:WAVelength:SWEep:DWELl<wsp><value>|MIN|MAX|DEF[NS|US|MS|S]
Description:
Sets the dwell time. Can only be used when sweep is stepped.
Cannot be set while a sweep is running.
Parameters:
The dwell time as a float value.
If you specify no units in your command, seconds are used as the default.
Also allowed are:
Response:
None
Example:
:sour0:wav:swe:dwel 500ms
Command:
:SOURce0:WAVelength:SWEep:DWELl?
Syntax:
:SOURce0:WAVelength:SWEep:DWELl?[<wsp>MIN|MAX|DEF]
Description:
Returns the dwell time. Can only be used when sweep is stepped.
Parameters:
None
Also allowed are:
Response:
The dwell time in seconds.
Example:
:sour0:wav:swe:dwel? -> +5.00000000E-001
Affects:
All N777xC tunable laser sources
70
MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the sum of, the minimum
programmable value and the maximum programmable value
MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the sum of, the minimum
programmable value and the maximum programmable value
Keysight N777xC Series Tunable Laser Family Programming Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

N7776cN7778cN7779c

Table of Contents