3
Instrument Setup and Status
Response:
The results for the individual slot events (a 16-bit unsigned integer value, where 0
Bit
8-16
7
6
5
4
3
2
1
0
Example:
:stat0:oper?
Command:
:STATus0:OPERation:CONDition?
Syntax:
:STATus0:OPERation:CONDition?
Description:
Returns the Operational Slot Status Condition Register of the laser module.
Parameters:
none
Response:
The results for the individual slot events (a 16-bit unsigned integer value, where 0
Bit
8-16
7
6
5
4
3
2
1
0
Example:
:stat0:oper:cond?
36
Mnemonic
Not used
Not used
Not used
Not used
Slot n: shutter has been opened
Slot n: Zeroing ongoing
Not used
Slot n: Coherence Control has been switched on
Slot n: Laser has been switched on
->
+0
Mnemonic
Not used
Not used
Not used
Not used
Shutter open
Zeroing ongoing
Not used
Coherence Control is switched on
Laser is switched on
->
+0
≤
≤
Keysight N777xC Series Tunable Laser Family Programming Guide
≤
value
65535):
Decimal Value
256
128
64
32
16
8
4
2
1
≤
value
65535):
Decimal Value
256
128
64
32
16
8
4
2
1