System:communicate:ethernet:ipaddress; System:communicate:ethernet:ipaddress:current; System:communicate:ethernet:macaddress - Keysight N777 C Series Programming Manual

Tunable laser source
Table of Contents

Advertisement

Parameters:
String (Up to four groups of up to 3 digits, groups separated by ".". Groups with leading
zeroes are interpreted as octal numbers.)
Response:
None
Example:
:syst:comm:eth:ipad "192.132.13.2"
Command:

:SYSTem:COMMunicate:ETHernet:IPADdress?

Syntax:
:SYSTem:COMMunicate:ETHernet:IPADdress?
Description:
Get the manually set IP address of the system.
Parameters:
None
Response:
String
Example:
:syst:comm:eth:ipad?
Command:

:SYSTem:COMMunicate:ETHernet:IPADdress:CURRent?

Syntax:
:SYSTem:COMMunicate:ETHernet:IPADdress:CURRent?
Description:
Get the current IP address of the instrument.
Parameters:
None
Response:
String
Example:
:syst:comm:eth:ipad:curr? -> "192.132.13.2"
Command:

:SYSTem:COMMunicate:ETHernet:MACaddress?

Syntax:
:SYSTem:COMMunicate:ETHernet:MACaddress?
Description:
Get the MAC address of the network adapter.
Parameters:
None
Response:
String (hexadecimal value).
Example:
:syst:comm:eth:mac? -> "00-07-E0-14-AE- 08"