Trigger:configuration - Keysight N777 C Series Programming Manual

Tunable laser source
Table of Contents

Advertisement

4
Measurement Operations & Settings
Response:
DISabled:
STFinished:
SWFinished:
SWSTarted:
Example:
:trig0:outp?
Affects:
All tunable laser modules
Command:

:TRIGger:CONFiguration

Syntax:
:TRIGger:CONFiguration
Description:
Sets the hardware trigger configuration with regard to Output and Input Trigger Connectors.
Parameters:
0 or DISabled:
1 or DEFault:
2 or PASSthrough:
3 or LOOPback:
Response:
none
Example:
:trig:conf dis
affects
All modules
Command:
:TRIGger:CONFiguration?
Syntax:
:TRIGger:CONFiguration?
Description:
Returns the hardware trigger configuration.
Parameters:
None
Parameters:
0 or DISabled:
1 or DEFault:
2 or PASSthrough:
3 or LOOPback:
Example:
:trig:conf?
affects
All modules
88
Never.
When a sweep step finishes.
When sweep cycle finishes.
When a sweep cycle starts.
->
dis
Trigger connectors are disabled.
The Input Trigger Connector is activated, the incoming trigger response for each
slot.
A trigger at the Input Trigger Connector generates a trigger at the Output Trigger
Connector automatically. No triggers from the instrument at the output.
The same as PASSthrough. This is included for compatibility reasons.
Trigger connectors are disabled.
The Input Trigger Connector is activated, the incoming trigger response for each slot.
A trigger at the Input Trigger Connector generates a trigger at the Output Trigger
Connector automatically. No triggers from the instrument at the output.
The same as PASSthrough. This is included for compatibility reasons.
->
DEF
Keysight N777xC Series Tunable Laser Family Programming Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

N7776cN7778cN7779c

Table of Contents