Integra CODMAN CERTAS Plus Manual page 17

Electronic tool kit
Hide thumbs Also See for CODMAN CERTAS Plus:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Guidance and manufacturer's declaration
The CERTAS Plus Electronic Tool Kit is intended for use in the
electromagnetic environment specified below. The customer or the user
of the Tool Kit should ensure that it is used in such an environment.
IMMUNITY test
Radiated RF
IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
Table 3
– electromagnetic immunity
IEC 60601
Compliance
test level
level
3 V/m
80 MHz to
3 V/m
2.7 GHz
Electromanetic
environment – guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the Tool Kit than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
d = 1.2 √ P 150 kHz to 80 MHz
d = 1.2 √ P 80 MHz to 800 MHz
d = 2.3 √ P 800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,
should be less than
a
the compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
16
EN – ENGLISH
b

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Codman 82-8852

Table of Contents