3. Use the middle buttons ( and ) to scroll up or down to select CALIBR ALL. Press
Rosemount CT5100
SEL.
Figure 10-27: Gas Service (CALIBR ALL)
The system will work through the zero and span calibration process for all readings.
The Please wait screen will be displayed. This takes approximately one minute per
gas depending on the purge time.
Figure 10-28: Calibration
The Finished screen displays when the zero and span calibration processes
complete.
Gas calibration