Emerson Rosemount CT5100 Reference Manual page 148

Continuous gas analyzer
Hide thumbs Also See for Rosemount CT5100:
Table of Contents

Advertisement

Repair and replacement
May 2020
3. To prevent scratching the mirrors do not apply any direct pressure to the mirrors.
4. Clean the mirrors using the lens tissue in the direction shown in
5. Reassemble the analysis cell.
148
Figure 12-13: Correct Cleaning Technique
A. Correct
B. Incorrect
CAUTION
Be careful to wipe across the mirror surface to prevent any residue marks appearing
on the mirror surface where the solvent has evaporated.
Figure 12-14: Cleaning the Mirrors
Reference Manual
00809-0100-4511
Figure
12-14.
Emerson.com/Rosemount

Advertisement

Table of Contents
loading