Replacing The O Detector - Emerson Rosemount CT5100 Reference Manual

Continuous gas analyzer
Hide thumbs Also See for Rosemount CT5100:
Table of Contents

Advertisement

Reference Manual
00809-0100-4511
4. Disconnect the wiring harness (H).
5. Remove and retain the four screws (B) nuts (G), plain washers (D) and spring
6. Remove and discard the unserviceable detector module (C).
7. Inspect the new detector module for damage.
8. Fit the new detector module in position on the detector bracket (E). Secure the
9. Connect the wiring harness (H) to the detector module (C).
10. Connect the trigger OUT harness (A) to the detector module (C).
11. Close the door of the electrical compartment and restart the analyzer.
12.15
Replacing the O
The O
determine which detector type is fitted: 1 or 2. Replace the O
the following steps.
12.15.1
Replace O
Procedure
1. Shut down the analyzer.
2. Open the door of the electrical compartment.
Rosemount CT5100
washers (F).
detector module with the four screws, nuts and washers retained in
and G).
detector is located on the underside of the optical compartment base plate. First
2
detector (type 1)
2
Refer to
Shutdown
procedure.
detector
2
Repair and replacement
May 2020
Step 5
(B, D, F,
detector identified using
2
135