Advertisement

Troubleshooting and diagnostics
May 2020
5. Examine the parameters on the display to determine if the suspect component has
6. Press PAGE to display the next page of parameters.
108
Figure 11-8: Laser 1 Diagnostics Menu
A. BACK button
B. PAGE button
C. HELP button
D. Parameters
The component diagnostic screen shows the parameters of the component
selected from the Diagnostics menu. In the example shown in
parameters of Laser 1 are shown. The parameters data displayed vary depending
upon which component was selected for diagnosis.
failed or is operting outside of its correct parameters. If you do not have the
necessary information to determine if the suspect component is faulty, contact
Emerson for advice and assistance.
If there are more parameters than can be displayed on the available screen space, a
PAGE message will appear in the top right corner of the screen as shown in
11-8, and the top right button will become the associated PAGE button.
In the example shown in
second page of parameter data for Laser 1.
Figure
11-8, pressing the PAGE button will display the
Reference Manual
00809-0100-4511
Figure
11-8, the
Figure
Emerson.com/Rosemount