Emerson Rosemount CT5100 Reference Manual page 119

Continuous gas analyzer
Hide thumbs Also See for Rosemount CT5100:
Table of Contents

Advertisement

Reference Manual
00809-0100-4511
4. Fully raise the upper part of the fuse holder (B) as shown in
5. Fit the replacement fuse into the fuse holder.
6. Close the door of the electrical compartment and restart the analyzer.
Rosemount CT5100
push out the old fuse (A).
Figure 12-3: Replacing the Fuse
A. Fuse
B. Fuse holder
WARNING
CORRECT FUSES
Failure to use the correct fuses may result in personal injury or death and/or
damage to persons and/or property.
Only replace fuses with fuses of the same type and rating.
Refer to
Table 12-1
and verify that the fuse is of the correct type and rating.
When lowering the upper part of the fuse holder, ensure that it "clicks" into place
when fully lowered.
Repair and replacement
May 2020
Figure 12-3
and then
119