Install
5.1

Site selection

The Rosemount CT5100 has a T3 temperature classification which specifies the maximum
surface temperature of the analyzer. Ensure that no combustible gas concentrations will
be present, whether on a continual or occasional basis, which have an ignition
temperature below the T-rating of the analyzer.
WARNING
FIRE AND EXPLOSION
Death, personal injury, and/or damage to persons and/or property may result if this is not
observed.
The analyzer's electrical compartment must not be opened unless the atmosphere in the
area is known to be below the ignitable concentration of combustible gases or materials,
or unless all equipment within the protected enclosure is de-energized.
WARNING
ELECTRIC SHOCK
The analyzer operates using mains voltage, which may cause death or serious injury to
personnel. Failure to observe this precaution will cause death, personal injury, and/or
damage to persons and/or property.
Ensure that the circuit breakers are set to Off and locked out and tagged out off before
removing the top cover or opening the front cover.
The analyzer is intended to be installed in a suitable Division 2 shelter to protect it from the
elements.
Provide sufficient space around the analyzer to allow the maintenance and servicing of the
unit.
Rosemount CT5100
Install
May 2020
41