system through the secretion canister and hose system to the patient-sided hose end. In this
way secretion and air can be sucked through the hose system or transported into the secretion
canister. A bacterial filter in the secretion canister protects the basic device from possible
contamination with bacteria and the suction of secretion.
The vacuum which is at the patient-side is measured with the measuring- and rinsing hose.
Furthermore after a defined period a valve opens and flushes secretion, coagulum and other
blockages from the hose to the secretion canister.
User profile: Doctor, medical staff
Patient groups: Patients of all ages with and without restrictions
Application organ: Thorax
Application time:
Secretion canister on the device: Short-term use on the patient (< 30 days).
Secretion canister for gravity drainage: For temporary use on the patient (< 60 minutes).
Area of application: The application site is the clinical area. The application of the device may
only be applied by healthcare professionals. The secretion canister and the drainage hose are
sterile and disposable, and can be applied in the sterile OT area.
Contraindications:
No application with thoracic drainage systems other than the ATMOS E / S 201 Thorax.
No separate use of the secretion canister and the hose system (i.e., without the basic device) for
short- or long-term gravity drainage (more than 60 minutes).
No application under emergency conditions and in the home care field which is not supervised
by healthcare professionals.
No suction of flammable, corrosive or explosive fluids / gases.
The product is: not active
Sterility: Secretion canister is sterile.
Single-use product / reprocessing: Secretion canister is disposable.

1.2.3 Intended use ATMOS S 201 Thorax hose system

Name: Hose system for the ATMOS S / E 201 Thorax and ATMOS C 051 Thorax
Main functions:
Transport of generated vacuum from the basic device to patient-sided hose end.
Suction of secretion and air through the secretion hose into the secretion canister.
Transport of the patient-sided vacuum to the vacuum sensor for detecting the current vacuum,
which is at the patient side.
Medical indications / application:
Creating the (natural) vacuum in the pleural cavity after a pneumothorax or a pleural effusion by
draining off air and secretion.
Drainage of secretion and air after a surgical opening of the thorax.
Specification of the main function: Generation of a vacuum from the thoracic drainage
system through the secretion canister and hose system to the patient-sided hose end. In this
way, secretion and air can be suctioned by means of the hose system or transported into the
secretion canister. The vacuum which is at the patient-side is measured with the measuring- and
rinsing hose. Furthermore after a defined period a valve opens and flushes secretion, coagulum
and other blockages from the hose to the secretion canister.
User profile: Doctor, medical staff
Patient groups: Patients of all ages with and without restrictions
Application organ: Thorax
Application time: Short-term use on the patient (< 30 days).
Application site: The application site is the clinical area. The application of the device may only
be applied by healthcare professionals. The secretion canister and the drainage hose are sterile
and disposable, and can be applied in the sterile OT area.