Contraindication Atmos ® S 201 Thorax; Warnings - Atmos S 201 Thorax Operating Instructions Manual

Hide thumbs Also See for S 201 Thorax:
Table of Contents

Advertisement

1.2.2 Intended use of the secretion canister for ATMOS
Name: Sterile secretion canister for the exclusive connection to the ATMOS
Main functions:
Forwarding of the created vacuum into the drainage hose and fi nally into the patient side drain.
Collection of the exhausted secretion.
Med. indications/ application: Secretion canister for the connection to a digital device for mobile thoracic drainage.
Specifi cation of the main function:
The vacuum generated by the pump evacuates air and secretion out of the drain. The secretion is collected in the secretion canister
whose fi lling quantity is 2 l. A bacterial fi lter prevents the potential contamination of the room and the device by bacteria. Additionally
it prevents the accidental intrusion of secretion into the pump head. A sterile tubing system provides the vacuum for the patient side
drain. A direct connection to the patient doesn't exist. With the support of the test measuring hose inside of the tubing system, which
is also provided with a fi lter, the actual vacuum on the drain will be measured. Furthermore the measuring hose is also responsible
for the fl ushing of the secretion hose with atmospheric air.
Application organ: Thorax
Application time: Short-term use on the patient (< 30 days).
Application site: Application environment is a clinical area. The application of the canister may only be applied by healthcare
professionals. The secretion canister and the drainage hose are sterile and disposable and can be applied in the sterile OT area.
Contraindications: No application of the canister with other low vacuum devices than ATMOS
The product is:
Sterility: Sterile
Single use product / reprocessing: Sterile single use product
1.3 Contraindications of ATMOS
In ambulances in service and in the home care fi eld which is not supervised by healthcare professionals.

1.4 Warnings:

Patient hose clamp must remain open during drainage operation.
The device is not for use while bathing, showering nor suitable for a hazardous explosive environment.
Do not use this device in MRI (Magnetic Resonance Imaging).
Do not dry the device with microwaves.
Chest hoses should not be clamped except when changing the canister. In the event of an airleak, clamping chest
hoses could lead to a tension pneumothorax.
The ATMOS
S 201 Thorax must remain in an upright position during use.
®
The patient should be regularly monitored according to internal hospital guidelines.
ATMOS
S 201 Thorax is a medical device that requires special safety precautions and must be installed and
®
placed in operation in accordance with the attached EMC information. Portable and mobile RF communication devices
(mobile telephones) can affect the drainage system.
A replacement device must always be available for patients for whom a breakdown of the device can lead to a critical situation.
Do not separate patient hose in-line connector prior to clamping off patient hose fi rst.
Do not keep patient hose clamp closed during drainage collection or patient transport.
Please note that when the set pressure is -50 mbar or higher this high pressure can cause pain or injuries to the patient.
4
□ active
X not active
S 201 Thorax
®
S 201 Thorax
®
E / S 201 Thorax.
®
E / S 201 Thorax.
®