Atmos S 201 Thorax Short Operating Instructions
Hide thumbs Also See for S 201 Thorax:

Advertisement

Quick Links

Short operating instructions ATMOS S 201 Thorax
Exchange of canister
click
click
Connection of
hose system
Display symbols
Battery status display /
charging indicator
Key-lock
Service display / Please call service
Service display / Please call service
Target vacuum
Uncoupling of the device may only be performed after pinching off the hose
and if a target value of 15 mbar is reached.
Exchange the canister at the latest when a fi lling level of 1900 ml is reached.
Briefl y press the I/O key, the pump stops.
Press the release button and swivel the canister system forward until the canister
can be easily removed from its guides.
Disconnect the canister from the hose system and the secretion channel. Seal the
secretion channel and the pop off valve with the protective caps.
There is no need to seal the connection to the device because the bacterial fi lter is
covered with a self-sealing coating.
Dispose of canister correctly.
Insert the new canister.
Start up the pump by pressing the I/O key.
Reopen the clamp at the suction hose.
1. Remove the sterile hose system from the sterile packaging.
2. Attach the hose connection with the bacterial fi lter to the lower connection (on the
canister) by clockwise rotation.
3. Connect the Luer-Lock connection (large diameter) to the upper connection (of the
canister) by clockwise rotation.
Setting up
Ensure a secure position of the device. Start the devices by pressing
the key.
Leakage test
Start the device by pressing the key
A leakage test is performed automatically.
To do this close the coupling on the end of the hose system.
If the leakage test was successful „Leakage test OK" appears
on the display. By pressing the key
on the display. By pressing the key
menu.
Only exit this menu when you wish to start the therapy.
Once the main screen appears the alarm system is activated.
When the SD card is already inserted during startup, then the display
indicator is on the left.
by pressing the
by pressing the
new recording.
Therapy stopped
Therapy in progress
Upcoming warning
Upcoming warning
suppressed
suppressed
Actual vacuum
Actual vacuum
Average value
Average value
Value changes every 10 seconds
Value changes every 10 seconds
Flow displayed as bubbles
Flow displayed as bubbles
Each additional dark colored bubble represents an additional flow.
Each additional dark colored bubble represents an additional flow.
From 1 to 9:
From 1 to 9:
0,25 l/min (blue)
0,25 l/min (blue)
from 10 to 12: 3,30 l/min (red)
from 10 to 12: 3,30 l/min (red)
Actual fl ow
Actual fl ow
average value of 2 seconds
average value of 2 seconds
you can go back to the main
you can go back to the main
button you can continue recording or start a
button you can continue recording or start a
Buttons
On/off switch
Increase target vacuum
Decrease target vacuum
Grafi c diagram of the therapy
Confi rm entry / reset

Advertisement

loading

Summary of Contents for Atmos S 201 Thorax

  • Page 1 Short operating instructions ATMOS S 201 Thorax Exchange of canister Uncoupling of the device may only be performed after pinching off the hose and if a target value of 15 mbar is reached. Exchange the canister at the latest when a fi lling level of 1900 ml is reached.
  • Page 2 Key lock After switching on the device the key lock is automatically activated if no key is operated within the preselected period (factory setting: 5 min). For manual activation of the key lock, please press keys For manual activation of the key lock, please press keys simultaneously until the simultaneously until the symbol is displayed on the screen.