4.0 Operation

4.1 Explanation of the display

Flow
Display of the actual fistula
ATTENTION: This is the current value. In
the case of a discontinuous fistula, the
value can decline temporarily to zero
although the fistula still exists.
Actual
vacuum
Display of
the actual
vacuum
value.
Target vacuum
Display of the
preadjusted
target vacuum
to which the
pump adjusts.
Starting / Stopping
therapy
Keylock activated
System information
Keylock manual
Flow displayed as bubbles
Each additional coloured bubble represents an additional
flow.
None:
0 - < 50 ml/min
Green:
50 - < 100 ml/min
Yellow:
100 - < 630 ml/min
Orange:
630 ml – 5.51 l/min
Red:
> 5.51 l/min to maximum.
Up to 1.00 l/min, the flow is displayed in ml/min.
Day/Night Mode
The ATMOS S 201 Thorax has a day/night
mode, which means the device adjusts
automatically to the light conditions in a
room.
Under low ambient light conditions,
the display has dark background illumi-
nation.
Therapy progress
Changeover to the
graphic diagram
User settings
Setting button
Increases the
target vacuum
Setting button
Decreases the
target vacuum
21