contamination with bacteria and the suction of secretion.
The vacuum which is at the patient-side is measured with the measuring- and rinsing hose.
Furthermore after a defined period a valve opens and flushes secretion, coagulum and other
blockages from the hose to the secretion canister.
User profile: Doctor, medical staff
Patient population: Patients of all age groups with and without restriction of any kind
Application organ: Thorax
Application time: Short-term use on the patient (< 30 days).
Application site: Application environment is a clinical area. The application of the device may
only be applied by healthcare professionals. The secretion canister and hose system are sterile
single use products, which can be applied in the sterile OT area.
Contraindications: No application with other thoracic drainage systems as the ATMOS
201 Thorax.
No separate application of secretion canister and the hose system (this means without basic
device) for gravity drainage.
No application in ambulances under emergency conditions and in the home care field which is
not supervised by healthcare professionals.
No suction of flammable, corrosive or explosive fluids / gases.
The product is: not active
Sterility: Secretion canister is sterile.
Single use product / reprocessing: Secretion canister is a single use product.
1.2.3 Intended use ATMOS
Name: Hose system for the ATMOS
Main functions:
Transport of generated vacuum from basic device to patient-sided tube end.
Suction of secretion and air through the secretion hose into the secretion canister.
Transport of the patient-sided vacuum to the vacuum sensor for detecting the current vacuum,
which is at the patient side.
Med. indications/ application:
Recovery of the (natural) vacuum in the pleural cavity e.g. after the occurrence of a
pneumothorax or a pleural effusion by drainage of air and secretion.
Drainage of secretion and air after a surgical opening of the thorax.
Specification of the main function: Generation of a vacuum from the thoracic drainage
system through the secretion canister and hose system to the patient-sided tube end. In this
way secretion and air can be sucked through the hose system or transported into the secretion
canister.
The vacuum which is at the patient-side is measured with the measuring- and rinsing hose.
Furthermore after a defined period a valve opens and flushes secretion, coagulum and other
blockages from the hose to the secretion canister.
User profile: Doctor, medical staff
Patient population: Patients of all age groups with and without restriction of any kind
Application organ: Thorax
Application time: Short-term use on the patient (< 30 days).
Application site: Application environment is a clinical area. The application of the device may
only be applied by healthcare professionals. The secretion canister and hose system are sterile
single use products, which can be applied in the sterile OT area.
Contraindications: No application with other thoracic drainage systems as the ATMOS
201 Thorax and ATMOS
No separate application of secretion canister and the hose system (this means without basic
6
S 201 Thorax hose system
®
S / E 201 Thorax and ATMOS
®
C 051 Thorax and corresponding secretion canister.
®
®
C 051 Thorax
®
®
E / S